Section 2

Introduction

Section 3

Unpacking and Setting Up

Fig. 2.1. Model EP-1 Econo Pump.

The Model EP- l Econo Pump is a two-channel, bi-directional, variable

speed peristaltic pump for low-pressure chromatography and general labo

ratory use. It offers a full range of features to facilitate ease of use as a stand

alone pump or as an integral part of the Econo System. As a stand-alone

pump, the Model EP-1 Econo Pump delivers flow rates from 0.l to 40 mumin

with the ability to self-calibrate the flow rate for 0.8, l.6, and 3.2 mm ID

tubing, displaying pump output in ml/min. The pump can be programmed to

control fraction collector parameters such as fraction size, void volume, and

total run volume. A membrane key panel with graphic icon displays allows

easy user interface. When coupled with the Model ES-1 Econo System

Controller, the Model EP-1 Econo Pump will control the gradient run-time,

gradient shape, and peak collection capabilities of the Econo System. When

coupled with the Model ES-1 Econo System Controller and Model EV- 1

Econo Buffer Selector, the Model EP-1 Econo Pump will control up to 5

buffers, automated sample injection, and method cycling.

3.1 Unpacking Instructions

Carefully remove the unit from the shipping box, lifting from the handle

on the back of the pump or from the bottom of the instrument. Remove the

plastic bag and inspect the instrument for any external damage. Parts includ

ed with the Model EP- I Econo Pump are illustrated in Figure 3.l. Check

off all parts against the supplied packing list.

Your Model EP- l Econo Pump was carefully tested at the factory and was

shipped in good working order. If any part is missing or damaged, contact Bio-

Rad Laboratories immediately. Refer to Figure 3.1 for the proper identifica-

tion designation of any missing or damaged part(s).

Also includes: Instruction Manual

Fig. 3.1. Parts supplied with the Model EP-1 Econo Pump.

3

2

Содержание EP-1 Econo Pump

Страница 1: ...Model EP 1 Econo Pump Instruction Manual Catalog Numbers 731 8140 731 8142 For Technical Service Call Your Local Bio Rad Office or in the U S Call 1 800 4BIORAD 1 800 424 6723...

Страница 2: ...8 5 l Tubing Selection 8 5 2 Installation of Precut Bio Rad Tubing 9 5 3 Installation of Uncut Tubing 11 Section 6 Operation 11 6 l Start Up Procedure I1 6 2 Verification of Pump Software Version 12...

Страница 3: ...Section 1 Safety Disconnect supply before servicing No user serviceable parts inside refer servicing to Bio Rad service personnel 1...

Страница 4: ...Econo Pump will control the gradient run time gradient shape and peak collection capabilities of the Econo System When coupled with the Model ES 1 Econo System Controller and Model EV 1 Econo Buffer...

Страница 5: ...ration Rotate the voltage selector through the window in the fuse drawer until the proper voltage shows 4 Reinsert the fuse drawer into the power entry module with the locking tab to the left The volt...

Страница 6: ...rter valve to collect or waste positions and total run volume The total run vol ume V may be used to program the pump to automatically shut off after a given volume has been delivered 4 2 Rear Panel F...

Страница 7: ...hree and will provide the most consistent flow rate over time Table 5 l shows approximate flow rate ranges with different tubing sizes Flow rates above 20 ml min may be obtained by plumbing two channe...

Страница 8: ...onto the pump head as described in Section 5 2 Warning If using tubing other than the type supplied by Bio Rad make sure the wall thickness is not greater than 1 0 mm Using tubing with a greater wall...

Страница 9: ...nt panel of the Model EP 1 Econo Pump The four digit LED display on the front panel of the Pump should dis play v2 01 or higher If your pump displays a number lower than 2 01 and you will be using the...

Страница 10: ...g prompting entry of fraction size Enter the fraction size in 0 l ml increments using the Arrow keys After frac tion size is selected press the Fraction Collector key once more to confirm the fraction...

Страница 11: ...display will show time elapsed The pump will stop automatically when the total run volume has been delivered 1 6 7 To abort the fraction collection scheme at any time dur ing the program press the Run...

Страница 12: ...o continue a run in FAIL progress press program run key Otherwise press the run stop key Pump will not run No power Verify that powercord is plugged no LEDs lit in and the power switch is turned on Fu...

Страница 13: ...than 15 milliamperes of current The following table describes the two rear panel mini DIN connector pinouts Table B Rear Panel Pinouts 21 Technical Specifications Number of channels 2 Flow rate range...

Страница 14: ...cks and tub ing connectors Cables 731 8261 Econo System Cable 1 8 pin mini DIN to DB 9 connector For connection of Model 2110 Econo Fraction Collector to the Econo Pump Econo UV Monitor or Econo Syste...

Отзывы: