User and service manual Stainer AUS124 from sn 101548475
Bio Optica Milano S.p.A.
via San Faustino 58
I-20134 Milano
Tel. +39 02.21.27.13.1
Fax Acquisti/39 02.21.54.155
Fax Assistenza/Contabilità +39 02.26.41.74.48
Fax V39 02.21.53.000
Revision 160804 of 04/08/16
Page 6 of 38
N. 1 dish (capacity 750 ml) in white thermoplastic material for baskets load/discharge in position 0.
N. 2 running water dishes (capacity 750 ml) in transparent thermoplastic material.
N. 2 independent electric valves which activate the water flow when there is the passage in running water.
Pressure regulator to regulate the water flow.
N. 2 lithium buffer batteries for emergency power supply in case of black-out. Instrument automatic restart when
returning the current.
Working features
General ignition switch.
Warning light of line voltage.
Emergency switch for stopping all the machine functions at any time.
Entire management of the instrument through use of single colour touch screen panel with intuitive graphics.
Possibility of visualizing steps of programme during the execution of the cycle.
Possibility of memorizing more than 50 different protocols of coloration, with maximum number of 30 passage
for each.
Possibility of programming for each station: time of immersion/oscillation and dripping from 00’ : 01’’ up to 50’ :
59’’.
Possibility of setting up in machine parameters time of pause between the stations from 00’ : 01’’ up to 50’ :
59’’. Pause time is set up by default at 00’ : 01’’.
Possibility of changing, during the execution of the programme, the speed of oscillation, dripping and speed of
rotation.
Possibility of programming the filter’s replacement.
Possibility of compiling the map of the reagents loaded in machine specifying: name of reagent and its duration
with automatic date of the insertion.
Visual and acoustic alert in case of end of use of the reagents (function - check reagents).
Page of visualization and description of alarms accessible also during the execution of the programme.
Possibility of stopping and restarting a programme (in case of possible controls, filling up of reagents, ecc.).
Visual and acoustic alarm in case of: cycle end, filter’s saturation, absence of current.
Visual and acoustic alarm in case of table out of position.
Visual and acoustic alarm for water discharge (malfunction of electric valve, obstruction discharge pipes, etc.).
Double reset of safety in case of accidental stop.
Possibility of visualizing the steps of the stations during the cycle.