12) PROGRAMMED MAINTENANCE AND CLEANING
Annual periodic maintenance
The annual periodic maintenance must be effected only by qualified and authorized staff. For information contact the
Technical Assistance Service.
Maintenance for the user
A proper standard of hygiene and the routine maintenance of the laboratory are necessary to keep the instrument in
good working conditions. The operator is responsible for the proper cleaning of the stainer. In particular, it is necessary
to control the state of water inlet/outlet pipes, draining central basin and running water dishes, which must be free of
mould, encrustations, etc. which might obstruct the water passage.
Important:
Do not use acid or abrasive cleaning products in order to avoid ruining the painted surface or other parts
(washers, etc.) and wear anti-cut gloves over or under the common gloves to avoid possible accidents. It is possible to
use bleach to clean the pipes.
13) MAIN SAFETY ELEMENTS
1) If necessary, it is possible to stop the machine movements by pressing the red emergency button placed in the
frontal side of the instrument. In order to restart the instrument, you must rotate the emergency button clockwise
and effect a reset of the drives (see paragraph 10) ALARMS).
2) The machine is provided with a sensor that detects the position of the pvc rotating table. If the table goes
accidentally out of position, the machine stops all the movements and signals the alarm “table out of position”.
3) A float sensor is positioned inside the discharge central sink. If there is a malfunction of the electric valves or an
obstruction of the discharge pipes that prevent the water from flowing, with consequent raising of the level, the
machine interrupts the water flow intervening on the electric valves. In this case the machine doesn’t stop but
signals the fault.
4) When the level returns to normal, automatically the electric valves start to work normally again.
14) MAIN COMPONENTS
CODE
DESCRIPTION
QUANTITY
40-400-050
Reagents basin
22
16-1621
Water inlet pipe
2,5 m
16-1622
Water outlet pipe
2,5 m
16-1623
Water basin/instrument black pipe
3 m
40-400-052
Slide holder basket
2
36-A01ES
Emergency button
1
36-POMAUT
Knob
1
36-VOLFIL
Threaded knob
2
36-PLC
Plc
1
36-TSCOLOR
Colour terminal touch screen
1
36-ALIM
Feeder
1
36-UPS
Ups
1
36-MOTOREPP Square fan
2
36-AZION.0
Drive
2
36-VENT
Fan
3
36-ELET
Electric valve + coil
2
36-SOLEN.0
Solenoid
2
36-MOLSOL
Solenoid spring
2