35
D. IF SERVICE IS REQUIRED
The MVP-10 Pediatric Ventilator should be returned to BMD, directly or through
your dealer, for any repair or service that may be required.
Maintenance scheduling of the Bio-Med Devices' MVP-10 Pneumatic Ventilator
is dependent upon frequency of use, condition of supply gases, and handling. It
is recommended that units operated on a regular basis be factory serviced
annually. For units used infrequently, a maximum of 3 years between factory
service is recommended. Calibration may be periodically verified by qualified
personnel following prescribed checkout procedures. The annual service
procedure should be performed by experienced personnel at the factory. Field
service beyond minor adjustment is not recommended due to the MVP-10's
unique pneumatic logic circuit.
Pack the instrument so that its meters, control knobs, and connections are
adequately protected. Ship pre-paid with your name & telephone on the
paperwork to:
Bio-Med Devices, Inc.
61 Soundview Road
Guilford, CT 06437