33 

3.  RELIEF VALVE

 

 

CYCLE/CPAP switch to CPAP 
PIP full CCW 
PEEP PRESSURE full CW 
INSPIRATORY TIME .5 sec 
EXPIRATORY TIME 1.0 sec 

 
Occlude patient circuit exhalation valve exhaust port and slowly increase flow to 
6 LPM.  Manometer should indicate between 65 and 80 cm pressure.  Check 
ADJUSTABLE RELIEF VALVE at patient outlet at rear of unit if pressure is less 
than 65 cm.  Increase flow to a total of 12 LPM and verify pressure is between 
65 and 80 cm.  Reduce flow to 4 LPM.  With exhalation valve exhaust port open 
to atmosphere, switch CYCLE/CPAP to CYCLE.  Manometer should reach a 
minimum of 65 cm during inspiratory phase. 

 
4.  MAXIMUM PRESSURE

 

 

CYCLE/CPAP switch to CYCLE 
PIP full CCW 
PEEP PRESSURE full CW 
TOTAL FLOW of 6 LPM 
INSPIRATORY TIME 2 sec 
EXPIRATORY TIME 1 sec 

 
Manometer should read 70 ±5 cm during inspiratory.  Turn PIP knob clockwise 
until pressure decreases.  Continue turning to OFF position (knob reference line 
at 6 o'clock position and resistance is felt).  Manometer fluctuation should be no 
more than 1 cm.  CAUTION: DO NOT FORCE VALVE. 
 

5.  PEEP PRESSURE 
 

CYCLE/CPAP switch to CYCLE 
PIP full CCW 
PEEP PRESSURE full CCW 
TOTAL FLOW of 6 LPM 
INSPIRATORY TIME .5 sec 
EXPIRATORY TIME 1.0 sec 

 
Manometer should read 18 ±3 cm during expiratory.  Turn PEEP PRESSURE 
knob clockwise until PEEP PRESSURE decreases.  Continue turning to OFF 
position  (knob  reference  line  at  6  o'clock  and  resistance  is  felt).    During 
expiratory,  manometer  should  read  less  than  1  cm  above  manometer  zero 
reading.  CAUTION:  DO NOT FORCE VALVE. 

Содержание MVP-10

Страница 1: ...O MED DEVICES MVP 10 NEONATAL PEDIATRIC VENTILATOR INSTRUCTION MANUAL Catalog 2050 Rev 042916 BIO MED DEVICES INC 61 SOUNDVIEW ROAD GUILFORD CT 06437 800 224 6633 FAX 203 458 0440 www biomeddevices co...

Страница 2: ......

Страница 3: ...INSTALLATION CONSIDERATIONS 15 A EQUIPMENT REQUIRED 15 B SUPPLY GAS 15 C MOUNTING BRACKET 15 D PATIENT CIRCUIT 15 E OPTIONAL EQUIPMENT 15 IV VENTILATOR SET UP 17 A CONNECTION OF GAS SUPPLY 17 B CONNE...

Страница 4: ...FREE STANDING AND DESIGNATED BY AN MR CONDITIONAL LABEL AND AN M AS A SUFFIX TO THE SERIAL NUMBER WHEN USING THE MVP 10 MRI VENTILATOR IN AN MRI ENVIRONMENT THE FOLLOWING PRECAUTIONS MUST BE TAKEN THE...

Страница 5: ...SN Serial Number REF Catalog Number Authorized Representative in the European Community The CE mark displayed on this product signifies that this device is in compliance with the European Medical Devi...

Страница 6: ...3...

Страница 7: ...Two 2 supply hoses 1 4 ID x 3 8 OD x 10 long with attached DISS fittings one hose for air and one for oxygen Bracket for mounting MVP 10 to column Instruction Manual and Warranty Card Infant Test Lung...

Страница 8: ...the blender bracket together and separate the two halves Wrap the two halves around the pole and loosely secure the bracket to the pole below the ventilator using the two screws Be sure the two pins...

Страница 9: ...and hoses supplied with the instrument Nor does the warranty cover abuse or misuse of the instrument or damage due to unauthorized servicing If service is required the instrument must be properly pack...

Страница 10: ...stance or system malfunction Breathing through the negative pressure relief valve requires a greatly increased work of breathing and only air is provided A situation in which the patient is breathing...

Страница 11: ...The high pressure alarm level should never be set above the lowest relief valve setting Do not continue using a ventilator which has been significantly impacted or abused Volume limited ventilators sh...

Страница 12: ...ssure limitation with or without Positive End Expiratory Pressure PEEP Intermittent Positive Pressure Ventilation IPPV Intermittent Mandatory Ventilation IMV Continuous Positive Airway Pressure CPAP C...

Страница 13: ...ned by INSP TIME and EXP TIME control settings These settings are approximate time reference points which are repeatable within 5 For greater accuracy an external monitoring device may be used The tid...

Страница 14: ...me limited The pressure developed in the patient s airway depends upon the total system patient compliance and the tidal volume being administered When the PIP control is set at a pressure above the p...

Страница 15: ...provided past the patient WYE allowing the patient to inhale from this flow while maintaining the airway at the pre set constant positive pressure b Continuous Flow or Constant Oxygen Administration C...

Страница 16: ...LOGIC CIRCUIT CYCLE CPAP SWITCH PEEP CPAP CONTROL EXP TIME CONTROL MAX PRESSURE CONTROL INSP TIME CONTROL EXPIRATION VALVE CONNECTION GAUGE CONNECTION PATIENT CONNECTION ADJUSTABLE RELIEF VALVE TWO WA...

Страница 17: ...system in liters minute CONNECTIONS Rear Panel 11 AIR Connection to pressurized source of clean dry oil free air 50 5 PSI Mates with female DISS air fitting 12 POWER OXYGEN Connection to pressurized s...

Страница 18: ...t PATIENT connection user adjustable to open between a maximum pressure of 80 cm H2O 10 cmH2O when turned fully clockwise to a minimum of 10 cmH2O 5 cmH2O when turned fully counterclockwise PEEP CPAP...

Страница 19: ...is installed externally and may be adjusted to vent pressures between 70 cm H2O 10 cmH2O and 10 cm H2O 5 cmH2O INSP TIME and EXP TIME control settings are calibrated at sea level and 20o C ATPD using...

Страница 20: ...gic and may render the time cycling inoperative The MVP 10 will operate with a supply pressure outside of the 50 5 PSI 345 34 5 kPa range but accuracy of settings may be impaired In no case should a s...

Страница 21: ...nd or low T connected to Gauge line A Humidifier or Pressure Alarm connection can affect system compliance Connecting hoses and water level in the Humidifier should be arranged so that system complian...

Страница 22: ...es without flow restricting devices e g flowmeter valve A 50 PSI 345 kPa supply must always be connected to the POWER OXYGEN fitting for proper operation of the pneumatic logic controls Hose fittings...

Страница 23: ...rovided by a heated flow through humidifier or artificial nose If a Humidifier is not used connect hose of patient circuit directly to ADJUSTABLE RELIEF VALVE e Connect patient port of WYE to a test l...

Страница 24: ...MULTIPLICATION FACTOR from CHART C or by use of slide rule Set flow rate controls for this total flow STEP 4 Set CYCLE CPAP switch in CYCLE position STEP 5 For oxygen concentration FIO2 other than air...

Страница 25: ...otal Flow VI 10 ml 0 5 sec Total Flow VI 20 ml sec 1 2 liters min From CHART A For 60 oxygen concentration Oxygen Flow VIO2 600 ml min 0 6 liters min Air Flow VAir 600 ml min 0 6 liters min STEP 6 Att...

Страница 26: ...pliance of the system by increasing total flow In the previous example 5 ml must be supplied for the compliance in the system as well as the prescribed 10 ml tidal volume for the patient Therefore adj...

Страница 27: ...lock ring to secure adjusting cap Adjust PIP control until the desired pressure limit is seen during inspiration Note that the PIP level is somewhat affected by the flow rate It should be set with th...

Страница 28: ...by setting flow so that the pressure gauge reading does not drop during patient inhalation STEP 4 For oxygen concentration other than air or 100 oxygen set by adjusting integral flow rate controls acc...

Страница 29: ...24...

Страница 30: ...3 5 3 2 30 9 8 1 46 8 5 5 2 5 53 11 6 5 4 5 31 4 3 5 0 5 46 9 5 6 5 3 53 13 5 8 5 5 31 8 7 1 46 11 7 5 3 5 53 15 9 6 31 11 5 10 1 5 46 12 5 8 5 4 53 16 9 5 6 5 32 3 5 3 0 5 46 14 9 5 4 5 54 6 3 5 2 5...

Страница 31: ...5 8 59 15 5 8 7 5 65 16 7 9 72 14 5 9 59 16 5 8 5 8 65 17 7 5 9 5 72 15 5 5 5 10 59 17 5 9 8 5 65 18 8 10 73 11 5 4 7 5 61 6 3 3 66 3 5 1 5 2 73 13 4 5 8 5 61 8 4 4 66 7 3 4 73 14 5 5 9 5 61 9 4 5 4 5...

Страница 32: ...5 5 80 10 2 5 7 5 93 11 1 10 80 12 3 9 94 6 5 0 5 6 80 13 5 3 5 10 94 7 0 5 6 5 81 8 5 2 6 5 95 7 5 0 5 7 81 10 5 2 5 8 95 8 0 5 7 5 81 12 5 3 9 5 95 8 5 0 5 8 82 4 5 1 3 5 96 9 0 5 8 5 82 6 5 1 5 5...

Страница 33: ...2 14 4 4 5 71 1 2 6 8 13 5 5 0 67 1 1 6 4 12 8 5 5 63 1 1 6 1 12 1 6 0 60 1 0 5 8 11 5 7 0 55 94 5 2 10 5 8 0 50 86 4 8 9 6 9 0 46 79 4 4 8 8 10 0 43 74 4 1 8 2 11 0 40 69 3 8 7 7 12 0 38 65 3 6 7 2...

Страница 34: ...180 25 2 4 Insp Time Exp Time Mult Factor 1 8 0 6 33 3 1 6 0 8 37 5 1 44 96 41 7 1 2 1 2 50 96 1 44 62 5 0 8 1 6 75 0 6 1 8 100 48 1 92 125 0 4 2 0 150 34 2 06 176 0 3 2 1 200 27 2 13 225 30 2 Insp Ti...

Страница 35: ...proach setting by turning knob in a clockwise direction Humidifier should have low compliance and water maintained at a high level to minimize compliance Expiration Valve must be positioned with diaph...

Страница 36: ...flow rate is changed the PIP and PEEP CPAP levels should be checked When a compressor is used as the power source steps should be taken to filter and dehumidify the room air before introducing it int...

Страница 37: ...and AIR fittings Hand tightening of these fittings is sufficient Do not over tighten with a wrench as the fittings could be damaged Never connect a water supply to these fittings Use only clean dry oi...

Страница 38: ...itch to CYCLE PIP full CCW PEEP PRESSURE full CW TOTAL FLOW of 6 LPM INSPIRATORY TIME 2 sec EXPIRATORY TIME 1 sec Manometer should read 70 5 cm during inspiratory Turn PIP knob clockwise until pressur...

Страница 39: ...conds C CALIBRATION The accuracy of the instrument s indicators and controls should be retained over its life as long as it has not been subject to abuse The calibration of the pressure gauge flow met...

Страница 40: ...For units used infrequently a maximum of 3 years between factory service is recommended Calibration may be periodically verified by qualified personnel following prescribed checkout procedures The ann...

Страница 41: ...A AUTHORIZED REPRESENTATIVE IN THE EUROPEAN COMMUNITY Bio Med Devices Official Agent in Europe is Medical Market I N T AB Sehlstedtsgatan 6 115 28 Stockholm Sweden Telephone 46 08 767 70 00 Fax 46 08...

Страница 42: ...the area of interest is in the same area or relatively close to the position of the device Therefore it may be necessary to optimize MR imaging parameters for the presence of this device Additional r...

Отзывы: