SEQUENTIAL
CIRCULATOR
MODEL 200
8
All Bio Compression Systems products are
U.S. BASED MANUFACTURER
Страница 1: ...INC Toll Free Phone 800 888 0908 Phone 201 939 0716 Fax 201 939 4503 E mail biosystems biocompression com www biocompression com SEQUENTIAL CIRCULATOR MODEL 2008 All Bio Compression Systems products are U S BASED MANUFACTURER ...
Страница 2: ...l confident that your pump and garment are backed by the best warranty and customer service in the industry By simply dialing our toll free number for customer service 800 888 0908 warranty repairs or adjustments will be performed in a timely manner with minimal inconvenience to you Should a warranty repair require an extended length of time if available a pump from our loaner inventory will be ma...
Страница 3: ...tended for use during sleep REFER TO DOCUMENTATION BEFORE USING AND SERVICING TYPE B APPLIED PART CLASS II PROTECTION INTENDED USE The Model SC 2008 Sequential Circulator is a manual not software driven 8 chamber sequential pneumatic compression device intended for either primary or adjunctive treat ment of primary or secondary lymphedema The device is also intended for additional or alternate tre...
Страница 4: ...ith protective end caps which envelope each end of the pump thereby suspending the pump on all four sides within the carton This packaging design prevents damage to the pump which would ordinarily be sustained when the carton is thrown or handled roughly by the carriers NOTE The carton and end caps should be saved for re use each time the pump is either transported or shipped When transporting for...
Страница 5: ...clockwise will enable the pressure adjust ment knob to again turn free without resistance Caution Care must be taken not to Over Tighten the inner locking knob as only minimal force is required to lock pressure setting in place Note Changing the inner knob to a 1 2 x 8 32 Allen head set screw will upgrade this feature to tamper proof 3 Pressure Gauge mmHg 4 Air Ports Numbered Ports 1 4 Ports 5 8 4...
Страница 6: ...UPPER EXTREMITY GARMENT 6 Latch Connector Bar To match up with air intake ports on face of pump 6 6 4 ...
Страница 7: ...S 120 V c a 60 Hz 0 5 A When servicing use only identical replacement parts Do not remove cover Refer to qualified service personnel Lors des reparations utilisez exclusivement des pieces de re change identiques No retirez pas le couvercle Consultez un tech niciaen qualifie WARNING REPLACE FUSE WHERE MARKED ADVERTISSEMENT REMETTRE LE FUSIILE A L ENDROIT INDIQUE Fuse rated 3 Amps Time Delay 250 VAC...
Страница 8: ...remove fuse After placing new fuse in cap slot push cap and fuse inward and turn clockwise to lock in place NOTE The outer safety fuse is the only item serviceable by some one other than a Bio Compression Systems technician at the factory Bio Compression Systems technicians have been trained specifically for the manufacturing and repair of all Bio Compression Systems products NOTE Having no electr...
Страница 9: ...be measuring 281 x 187 with a Tolerance of 005 OPERATING INSTRUCTIONS Having familiarized yourself with controls and features of this equipment you are now ready to begin your treatment according to your physician s prescribed course of therapy 1 Make sure that your circulator is plugged into a safe properly secured 110 V AC outlet 2 Place unit on a sturdy table or other type of surface closed to ...
Страница 10: ... knob as it only takes slight pressure to lock the pressure adjustment knob in place NOTE You may have to repeat this procedure several times given the short cycle time for each chamber and since you can only set the pressure on port number one the high est pressure of the eight chambers After you have set the desired pressure shut the pump off until you are ready to use it and have your garment s...
Страница 11: ...he LATCH CONNECTOR BAR 6 to the remaining AIR SUP PLY PORTS 4 if using more than one garment 6 Unzip sleeve garment gently down to bottom stop zipper does not separate Place garment onto arm or leg and zip up to top of garment 7 Press PUMP SWITCH 1 up to PUMP ON position Allow two to three complete inflation resting cycles before the garment reaches its pre set therapeutic pressure At this point y...
Страница 12: ...l outward to remove garment from pump 2 Grasp four tubes in hand near LATCH CONNECTOR BAR 6 and pull plug outward away from pump to remove 3 First gently bend your arm or leg depending where garment is located to release partial air from chambers 4 Continue to assist in the evacuation of air from garment working from top to bottom 5 Once the garment feels loose enough you can unzip the garment all...
Страница 13: ... Garments 1 year from date of purchase invoice 11 REPAIR SERVICE 1 800 888 0908 BIO COMPRESSION SYSTEMS INC 120 West Commercial Ave Moonachie NJ 07074 www biocompression com NOTES _____________________________________________ _____________________________________________ Serial Number _______________________________________ Date Purchased ______________________________________ Local Representative...
Страница 14: ...ace of garment may require additional washing by hand with a clean towel while submerged Avoid using any abrasive materials such as scrubbing pads or chemicals that could cause damage to the exterior surface of garment 7 Re Submerge garment for 30 minutes with exception of tubing connectors in solution consisting of 1 cup of bleach per 1 gallon of warm tap water again agitating garment every 5 to ...
Страница 15: ... A Other bio compression systems products Also available Sequential Circulator Model 2004 Sequential Circulator Model 2004 FC Fast Cycle Sequential Circulator Model 3004 Sequential Circulator Model 3004 FC Fast Cycle Sequential circulator model 3008 The Biocryo system Multi Flo DVT Combo Prophylasix System BioArterial Plus Arterial Blood Flow Enhancement System Compression therapy garments Includi...