40
6 Set up with the display and adjustment module
NCR-84 Two-wire 4 … 20 mA/HART
51141-EN-181127
Setup:
Settings, e.g., for measurement loop name, units, application,
adjustment, signal output
Display:
Settings, e.g., for language, measured value display, lighting
Diagnosis:
Information, for example, on device status, peak value,
simulation, echo curve
Additional adjustments:
Date/Time, reset, copy function, scaling,
current output, false signal suppression, linearization, HART mode,
special parameters
Info:
Instrument name, hardware and software version, calibration
date, instrument features
In the main menu item "
Setup
", the individual submenu items
should be selected one after the other and provided with the correct
parameters to ensure optimum adjustment of the measurement. The
procedure is described in the following.
Here you can assign a suitable measurement loop name. Push the
"
OK
" key to start the editing. With the "
+
" key you change the sign and
with the "
-
>" key you jump to the next position.
You can enter names with max. 19 characters. The character set
comprises:
•
Capital letters from A … Z
•
Numbers from 0 … 9
•
Special char - / _ blanks
In this menu item you select the distance unit and the temperature
unit.
For the distance units you can choose between m, in and ft and for
the temperature units °C, °F and K.
This menu item allows you to adapt the sensor to the measuring
conditions.
Setup - Measurement
loop name
Setup - Units
Setup - Application
925-0373 Rev A