GWR-2000 Four-wire 4-20 mA/HART
925-0335 Rev B
9 Diagnosis, asset management and service
49
Code
Text mes-
sage
Cause
Rectification
DevSpec State in CMD 48
F261
Error in the
instrument
settings
- Error during setup
- Error when carrying
out a reset
- False signal suppres-
sion faulty
- Carry out a reset
- Repeat setup
Bit 9 of Byte 0 … 5
F264
Installation/
Setup error
- Error during setup
- Check or correct installation
and/or parameter adjustment
- Check probe length
Bit 10 of Byte 0 … 5
F265
Measurement
function dis-
turbed
- Sensor no longer car-
ries out a measure-
ment
- Carry out a reset
- Disconnect operating voltage
briefly
Bit 11 of Byte 0 … 5
F267
No execut-
able sensor
software
- Sensor cannot start
- Exchanging the electronics
- Send instrument for repair
No communication possi-
ble
The following table shows the error codes and text messages in the status message "
Function
check
" and provides information on causes as well as corrective measures.
Code
Text mes-
sage
Cause
Rectification
DevSpec State in CMD 48
C700
Simulation
active
- A simulation is active
- Finish simulation
- Wait for the automatic end af-
ter 60 mins.
"Simulation Active" in
"Standardized Status 0"
The following table shows the error codes and text messages in the status message "
Out of
specification
" and provides information on causes as well as corrective measures.
Code
Text mes-
sage
Cause
Rectification
DevSpec State in CMD 48
S600
Impermissible
electronics
temperature
- Temperature of the
processing electron-
ics in the nonspecified
section
- Check ambient temperature
- Isolate electronics
- Use instrument with higher
temperature range
Bit 8 of Byte 14 …24
S601
Overfilling
- Level echo in the close
range not available
- Reduce level
- 100 % adjustment: Increase
value
- Check mounting socket
- Remove possible interfering
signals in the close range
- Use coaxial probe
Bit 9 of Byte 14 …24