MK (E2.1) 10/2014
page 32/73
8.
Manual Mode
In Manual Mode (HAND) you can enter a temperature set-point, the fan speed (0% to 100%), and the
switching-state of up to 8 operation lines. Operation line 1 is used to control the bedew protection (chap.
10). The other operation lines are non-functional. All settings remain valid for Manual Mode (HAND) until
the next manual change, if the unit had been turned off or in case of toggling to Idle Mode or Program
Mode (AUTO).
8.1 Set-point entry
W
X
CONFIG
HAND
PGM
VIEW->
08:43:55 15.12.13
CONFIG
CONFIG
CONFIG
FAN
0
%
Idle Mode
No heating or cooling function.
The fan is off (factory setting)
Toggling to Manual Mode
Change of page
08:43:55 15.03.13
HAND
HAND
Hand-Mode
TEMP
SPEED
Steuerkontakte
+29.3 °C
+80.0 %
Hand-Mode
TEMP
FAN SPEED
Operating contacts
+40.0 °C
+50.0 %
PGM
CONFIG
W
VIEW->
EXIT
Toggling between temperature set-point, fan speed, and operation line 1 (bedew protection).
08:43:55 15.12.13
HAND
08:43:55 15.12.13
Operating contacts
Contact 1
Contact 2
Contact 3
Contact 4
Contact 5
Contact 6
Contact 7
Contact 8
On
Off
Aus
Aus
Aus
Aus
Aus
Off
HAND
Contact 1
Off
On
Hand-Mode
TEMP
FAN
Oper
+40.0 °C
+80.0 %
TEMP
+0060.0
°C
EXIT
Arrow buttons to select the operation lines
(Operation line 1 = Bedew protection)
Arrow buttons to
enter the value
Button to move the decimal point
Unlock the keyboard locking (option, chap. 14.3) via the key switch to enter the set-point.