FDL (E2) 03/2011
page 19/68
2.2
Control panel
(6) (5) (4) (3)
(2) (2b) (2a) (1)
Figure 5: Control panel of standard unit
(1)
Main power switch ON/OFF
(2)
Temperature safety device class 2
(2a) Red pilot lamp for temperature safety device class 2
(2b) RESET key for temperature safety device
(3)
Red indicator light, loss of technical ventilation
(4)
RESET key for the heating cut-out due to failure of technical ventilation
(5)
Temperature program controller RD3
(6)
Solvent curve: Highest permissible amount of solvent G
total
[g] as a function of the drying tempera-
ture
2.3
Solvent curve FDL 115
0,00
5,00
10,00
15,00
20,00
25,00
30,00
35,00
50
60
70
80
90 100 11
0
120 130 140 15
0
160 170 18
0
190 200 210 22
0
230 240 25
0
26
0
270 280 29
0
30
0
Temperature [°C]
Gtota
l [g]
Figure 6: Solvent curve FDL 115
The diagram shows the highest permissible solvent quantity G
total
[g] in the steam room in correspon-
dence to the drying temperature. This is based on the calculation acc. to EN 1539:2009 considering the
unit specific data, an assumed molecular weight of the solvent of 100g/Mol, and a lower explosion limit of
40g/m
3
at 20 °C /
68 °F
and at 760 Torr (1013 hPa) (assumptions for unknown solvents acc. to EN
1539:2009).