10.3 Simple troubleshooting
Defects and shortcomings can compromise the operational safety of the chamber and can lead to risks and
damage to equipment and persons. If there are is a technical fault or shortcoming, take the chamber out of
operation and inform BINDER Service. If you are not sure whether there is a technical fault, proceed
according to the following list. If you cannot clearly identify an error or there is a technical fault, please
contact BINDER Service.
Only qualified service personnel authorized by BINDER must perform repair.
Repaired chambers must comply with the BINDER quality standards.
Fault description
Possible cause
Required measures
General
Chamber without function.
No power supply.
Check connection to power
supply.
Check whether the chamber is
turned on at the main power
switch.
Wrong voltage.
Check power supply for correct
voltage (chap. 4.1).
Chamber fuse has responded.
Contact BINDER service.
Controller defective.
Temperature
Set-point temperature is not
reached after specified time.
Chamber door not properly
closed.
Completely close chamber door.
Door gasket defective.
Replace door gasket,
Controller not adjusted..
Calibrate and adjust controller.
The fan doesn’t turn or turns too
slowly.
Fan speed set too low
Set the fan speed to 100%.
Fan defective.
Contact BINDER service.
Chamber heating permanently,
set-point not held.
Controller defective.
Contact BINDER service.
Pt 100 sensor defective.
Semiconductor relay defective
Controller not adjusted..
Calibrate and adjust controller.
Chamber doesn’t heat up.
Red heating control light in the
display is lit.
Heating element defective.
Contact BINDER service.
Semiconductor relay defective.
Chamber doesn’t heat up.
Red heating control light in the
display is not lit.
Controller display working.
Timer has run off.
Program the timer or change to
time function Continuous operation
(chap. 6.3)
Semiconductor relay defective.
Contact BINDER service.
Controller defective.