Quick Start Manual Rev 1.33
Page 20
003X-608-133
GROUP
(GRP)
ITEM
(ITM)
SELECT
(SEL)
EDIT
(EDT)
Underline = Defaults
OK
OPTION
(ctd)
Z.BAND
⊗
Zero 'Dead' Band
<SEL>
changes position,
<EDT>
changes digit.
Default =
0
Accept
R.ENTRY
⊗
Enable Rear Pushbutton
for Full Setup Entry
OFF, ON
Accept
CAL
ZERO
⊗
Zero Calibration Routine
Set between zero and full
scale
(Current weight displays)
<SEL>
key to start. Current weight displays.
Remove all weight.
<SEL>, <EDT>
or
<OK>
starts routine (
Z.in P
displays).
<ITM>
key to exit,
<SEL>, <EDT>
or
<OK>
to repeat routine.
SPAN
⊗
Span Calibration Routine
Set between zero and full
scale
(Current weight displays)
<SEL>
key to start
<EDT>
key to select clear linearisation
points (
CLr.L y
displays).
or not (
CLr.L n
displays).
<OK>
key to select alternative.
Current weight displays.
Add test weight.
<SEL>
or
<OK>
to show calibration weight value.
Set correct weight.
<SEL>
changes position,
<EDT>
changes
digit.
<ITM>
or
<OK>
starts routine (
S.in P
displays).
<ITM>
key to exit,
<SEL>, <EDT>
or
<OK>
to re-edit
calibration weight and repeat routine.
Содержание B-TEK Scales T503W