RESPIFIT S
22
6 MAINTENANCE
The inspiratory muscle training device RESPIFIT S is designed as a low-
maintenance device.
For long-term maintenance of quality and functional safety, please
observe the following points:
The handheld patient module must only be used repeatedly by a single
patient for reasons of hygiene. The cleaning directions, as set out in the
instructions for use, must be meticulously observed in order to achieve
repeated use of the handheld patient module (for cleaning instructions
see Section 5).
Avoid exerting force on the training device or its accessories.
If the device is dropped, damaged due to force, or functions in a way
other than described in the instructions for use, stop using the system
and return it to the service centre.
6.1 Troubleshooting
Errors
Possible cause
Solution
Display is
blank.
No power supply
Insert the mains plug.
If the error persists, send the
device for servicing.
“DATAKEY “
Datakey is not
inserted or
datakey is inserted
incorrectly or
datakey is
defective.
Insert
datakey.
If
the
error
persists, the datakey may be
incorrectly
inserted
or
faulty.
If the datakey is faulty, request a
new one from the doctor /
therapist.
6.2 Periodic Inspections
The periodic technical safety inspections (according to EN 62353, for
example) must be carried out on the therapy device at least every 24
months. Contact the treating physician / therapist, the clinic or the retailer
to arrange this. The periodic inspection must be carried out by trained
and knowledgeable persons who have practical experience and are
qualified to carry out such technical safety inspections.
•
The safety labels on the device and its accessories must be easily
legible.