Instructions for Use – BW 685 / BW 685 S BIEGLER
Supplementary safety instructions for the accessory tube warmer Biegler
Tubeflow BW 685 S:
•
The Biegler Tubeflow may only be clamped to the Biegler blood
warmer BW 685 S with the fixing clamp according to the diagram in the
chapter "Initial operation".
•
Do not kink, cover, warm or cool the Biegler Tubeflow. Do not cover
the Tubeflow with cloths or dressing material. Do not lead under blankets,
warm air blankets, etc. Do not expose to direct sunlight or heat radiation.
Lead the Tubeflow through the air unencumbered. Do not lead into or
through incubators. Overwarming of the warmed liquid or burns to the
patient cannot to be excluded in the case of non-compliance.
•
Do not shorten or damage the Biegler Tubeflow.
The BW 685 / BW 685 S may not be used if:
•
the housing is damaged or one of the front film layers becomes detached.
•
the device has been exposed to a hard physical shock (e.g. dropped, hit or
shaken)
•
the device has been immersed in water
•
the device has triggered a high temperature alarm that was not caused by
external factors.
•
the mains power cord or plug is damaged
•
the device has given somebody an electric shock
•
the fixing clamps are damaged and no longer assure safe clamping to the
infusion stand.
Should a malfunction be evident, suitable warning signs should be attached
to the device to ensure that it cannot be used before necessary service and
repair work has been carried out.