Pre-set selector:
This three way switch allows
selection of either variable speed (VAR), or one of
the two pre-set speeds (P1 and P2).
Speed controller:
This knob controls the paddle
speed in conjunction with the press to set switch.
Turning it to the right increases the speed,
turning it to the left decreases the speed.
Digital display:
This screen displays the paddle
speed in rpm.
Operation
(see Fig 1 & 2)
Place the unit on a firm, level surface and
connect to the mains supply. Leave sufficient
space to access the side IEC socket.
Turn the mains and lamp switch on and observe
the fluorescent lamp illuminate. Use the blue
knob to move the stirrer paddles up or down as
required.
To position a beaker, lift the paddle up to its
maximum height using the paddle lift and place
the vessel over the illuminated base. Once the
samples are in position, lower the paddles into
the beakers to the required depth.
Selecting a variable speed
Set the Pre-set Selector to the VAR (variable)
position. Hold down the 'Press-to-set' switch and
turn the Speed Controller until the desired speed
in rpm appears on the digital display.
Selecting a pre-set speed
Set the Pre-set Selector to the P1 position. To
change the fixed speed, insert the blade of a
small screwdriver and turn the screw control
clockwise or anticlockwise until the desired speed
in rpm appears on the digital display. Now every
time the Pre-set Selector is put to the P1 position,
the stirring paddles will rotate at that speed.
If a second pre-set speed is required, set the
Pre-set Selector to the P2 position and repeat the
procedure.
Maintenance & Servicing
WARNING:
Ensure the unit is disconnected
from the mains electricity supply before
attempting maintenance or servicing.
Periodically clean the instrument using a damp
cloth and mild detergent solution. Do not use
harsh or abrasive cleaning agents.
Any repairs or replacement of parts MUST be
undertaken by suitably qualified personnel
.
The following spares and accessories are available
from your laboratory dealer.
Description
Catalogue
Number
Fuse
ESM3C15(S)UL
Mains On/Off
ESM3C21(S)
Paddle
CSW124
Light tube
ESW501(S)
Motor
MSW501(S)
For a comprehensive list of parts required by
service engineers conducting internal repairs and
a service manual, please contact the Service
Department at Bibby Scientific Ltd. quoting both
the model and serial number.
Only spare parts supplied or specified by Bibby
Scientific Ltd. or its agents should be used. Fitting
of non-approved parts may affect the
performance and safety features designed into
the instrument.
If in any doubt, please contact the Service
Department of Bibby Scientific Ltd. or the point
of sale.