Thank you for purchasing this Bibby
Scientific product. To get the best
performance from the equipment, and for
your personal safety, please read these
instructions carefully before use.
This equipment is designed to operate under the
following conditions:-
❖
For indoor use only
❖
Use in a well ventilated area
❖
Ambient temperature range +5°C to +40°C
❖
Altitude to 2000m
❖
Relative humidity not exceeding 80%
❖
Mains supply fluctuations not exceeding 10%
❖
Overvoltage category II IEC60364-4-443
❖
Pollution degree 2 IEC664
If the equipment is not used in the manner
described in this manual the protection provided
by the equipment may be impaired.
Electrical Installation
THIS EQUIPMENT MUST BE EARTHED
Before connection please ensure that the
line supply corresponds to that stated on the
rating label.
This unit requires a supply rated at 230V,
50-60Hz, single phase. Power consumption is
100W.
There is an IEC socket at the rear of the
instrument for connection to the mains supply.
The unit is supplied with two mains leads fitted
with IEC plugs for connection to the instrument.
One has a U.K. 3 pin plug and the other has a 2
pin “Shuko” plug for connection to the mains
supply. Choose the lead appropriate for your
electrical installation and discard the other.
Should neither lead be suitable, take the lead
with the U.K. plug and replace the plug with a
suitable alternative. See the enclosed instruction
sheet for advice on how to carry out this
procedure.
Should the mains lead require replacement a
cable of 1mm
2
of harmonised code H05W-F
connected to an IEC 320 plug should be used.
1
N.B. The UK mains lead is protected by a 10A
fuse mounted in the plug top.
The mains lead should be connected to the
instrument BEFORE connection to the mains
supply.
IF IN DOUBT CONSULT A QUALIFIED
ELECTRICIAN
Safety Advice Before Use
❖
Never move or carry the unit when in use or
connected to the mains electricity supply.
❖
Never move or carry the unit while beakers
are in place.
❖
The SW5 may be carried by the handle on the
top of the unit.
General Description
The Stuart SW5 Flocculator is designed to ensure
repeatable conditions between samples during
water and sewage flocculation tests.
The speed control has electronic feedback to
accurately maintain the rotational speed of the
stainless steel paddles. A digital display clearly
indicates the speed, which is variable between 25
and 250 rpm. Alternatively, the use of two
adjustable pre-sets allows repetition of
pre-selected speeds at the flick of a switch.
All paddles can be easily lifted out of the samples
when required. Observation and monitoring of
samples is assisted by diffused base illumination
and a black background.
Controls
(see previous page)
Figure 2 shows the controls for the SW5.
Mains on/off:
This is a rocker type switch which
illuminates when the unit is on.
Lamp on/off:
This is a rocker type switch which
illuminates when the unit is on.