15
3 The instrument status will change to
“Heating” and the command toolbar options
will change.
4 The SMP40 will quickly heat the sample block
to the set start temperature.
5 When the sample block reaches the start
temperature the SMP40 status will change to
“Stabilising” and the temperature of the
sample block will remain constant for 60
seconds.
6 After the instrument has stabilised at the start
temperature the status will change
automatically to “Analysing” and the
instrument will begin to heat the sample
capillaries at the specified ramp rate.
7 The Capillary tubes melt status is shown to
indicate the melt determinations progress.
The Capillary tubes melt status is shown for
Information Only.