Addendum sheet M7917 issue 2.3                            Page 1 of 3 

 

 

Method for connecting a Heating Tape / cord to an MC5 / MC227/ 
MC228x1 Controller.  
 
ATTENTION: INSTALLATION MUST BE UNDERTAKEN BY A 
COMPETENT ELECTRICIAN. 
 

To connect a Heating Tape to an MC controller you will require a connecting lead set. This 
sheet details a process for assembling a suitable lead set.  
 

Carefully examine your heating tape. Look at the terminal block and identify if you 
have a two lead or three lead type tape / cord.  
 

                      

 
               

 

                

Figure 1                                                           Figure 2                 

 
 
 
 
 
 
 
 
                Figure 3                                                         Figure 4 
 

Select suitable connection cable type giving due consideration to current, 
temperature and insulation requirements.  
Note:

 

The wire gauge should 

NOT

 be less than 

16/0.2mm.

 

 

Take IEC plug M7131. Using a screw driver, unscrew and remove the retention screw 
from the centre of the housing. Remove the top of the IEC plug. Reserve plug top and 
screw for use later. 
 

 

 