47
Care and Maintenance
Verification of the Charger’s Power
Source
If the light does not illuminate when there are
batteries correctly installed in the Voyager Duo, try
the following:
1) Make sure that the power cord is correctly
plugged into the charger and in the wall AC
outlet, and that the green light on the clip-on
charging station is on.
2) Make sure that there is contact between the
contact blades of the lift and the contact plates of
the charging station.
3) Check the power of the AC outlet on the wall.
4) If the charger’s green light does not light up,
contact your local BHM Medical representative
for assistance.
Sling Inspection and
Care
For maximum resident safety and hygiene, read the
instructions found in the
Loop Sling Instructions for
Use
that comes with the sling.
Annual Inspection
The Voyager Duo and its accessories must be
inspected annually by a qualified technician.
CAUTION: Do not attempt to use a battery
that was not supplied by BHM Medical.
BHM Medical batteries are specially
designed for BHM Medical charging
systems. Attempting to use an
unauthorized battery may seriously
damage the lift and/or the charger.
WARNING: The Voyager Duo and
accessories must be serviced every 12
months as a minimum requirement. Do not
attempt to do the inspection unless you are
certified to do so.
Maintenance Requirements
The Voyager Duo is equipped with an electronic
monitor that causes a red light to flash when a
maintenance inspection is necessary. Arranging for
scheduled inspections ensures the durability of the
unit and the security of the resident and user.
Содержание Voyager V5 Duo