10
Note:
While the jaw set code is flashing use the button to choose the size and the button to
select
c
for crimper or
d
for decapper.
Figure 5.
When the selection
is complete, press and hold
the go-button for 2 seconds
to confirm the selection.
Figure 6.
The display will
show
On
momentarily and
then enter operating mode.
Adjusting the tool for crimping jaws
The electronic crimping tools must be adjusted for the vials, caps, and seals that will be used. The
and adjustment buttons on the top of the crimping tool set a stop position for the motor that
drives the tool. Pressing either button one time displays the current setting. Pressing again will change
the setting.
The last setting for a given jaw set is reloaded when it is selected for use.
Figure 7.
The numerical
setting of the crimping tool
sets a stop position that
determines the amount
of compression of the
cap and is very accurate.
There may be some
drifting in the setting over