LPR Meter 

Handbook 

Page 2 

 

 

 

 

2. Major 

Components 

 

 

 

 

 

Handheld Meter 

 

 

 

 

 
 
 

 

 

 

 

 
 

Bigfoot 

 

 
 

 
 

 

Reference Electrode 

 

 
 

 

Reinforcing Bar Connection 

 

 

 
 
The following section outlines the major components of the 

system with a detailed explanation to there use. 

 

 

 
The handheld meter is the major component within the 

system and carries out the Linear Polarisation test after 

pressing a series of buttons.   

 

The meter provides the following functions: 
 

 

Date & time 

 

Data logging up to 512 Readings 

 

PC Download Capability 

 

PT100 temperature Measurement 

 

LPR ID Reader (permanent probes only) 

 

 

 

To carry out the corrosion rate measurement the big foot is 

used to record the Corrosion potential (Ecorr) utilising the 

reference electrode and pass a small current through the 
integral Auxiliary electrode (AE). 
 

 

 

The reference electrode is attached to the Big foot for 

recording the corrosion potential from the reinforcing steel. 
 

 

 

In carrying out the corrosion rate measurement the 

handheld meter must be provided with a connection to the 
reinforcing steel.  This is carried out by either a single 
connection (WE) if less than 10metres from the meter or by 

two connections (WE & SENSE)  for distances up to 30 

metres. 

 

NOTE:  No testing should be carried out in excess of 30 

metres and additional reinforcing steel connections are to be 
made. 

 

When using a single connection to the reinforcing steel the 

supplied link must be installed between the WE and SENSE 
terminals on the meter. 

 
 

 

 

 

 
 

 

Содержание BF225

Страница 1: ...LPR Meter Handbook...

Страница 2: ...CONTENTS Section Number Page Number One Introduction Page 1 Two Major Components Page 2 Three Operating Instructions Page 3 Four Corrosion Rate Measurement Operation Page 6 Five LPR Standard Page 9...

Страница 3: ...mplex nature of corrosion rates by carrying out all of the calculations internally The meter is supplied in a IP65 enclosure and is suitable for external use The handheld meter equipment consist of th...

Страница 4: ...d to record the Corrosion potential Ecorr utilising the reference electrode and pass a small current through the integral Auxiliary electrode AE The reference electrode is attached to the Big foot for...

Страница 5: ...above diagram provides the connections required to carry out the corrosion rate measurement Note that either two connections are provided to the reinforcement or a single connection is acceptable if l...

Страница 6: ...is supplied with a separate manual and is not provided as standard please consult manufacture for data logging software download Black is reference electrode RE Blue is sense short to Green if not use...

Страница 7: ...nd corrosion rate are displayed Select EITHER button Meter confirms that data was stored Select GREEN button for next reading and repeat RED button starts cycling through displays of stored data as Da...

Страница 8: ...qual to the polarisation current times the range resistance The second output is proportional to the absolute value of the polarisation current at the end of the polarisation period It is generated by...

Страница 9: ...autoranging The output of the current measuring amplifier always reflects the current going to the test electrode The DAC2 output is set when the GO line is released to high to reflect the cell curre...

Страница 10: ...ectrode Hence E is 0 02V substituting in the above and solving gives 87 7 1 2 10 645 9 85 55 60 60 24 365 02 0 4 area B I rate n penetratio applied year m area I rate n penetratio applied 35 lapplied...

Страница 11: ...LPR Meter Handbook Page 9 5 LPR Standard...

Отзывы: