BLUEDENT SMART cordless 

Dental curing light 

 

12 

Remove the light guide and looking through it at strong light: if there are darkened 
areas over 10% of the area, replace it with a new one. 

OK 

 

 

Defective 

 

 
Lightguide 

should be sterilized by autoclave at 134° C for 5 min.

 

The same must be done for protection cap. 
The protective filter is wiped only with disinfectant. 

2. Cleaning of the unit.  
For disinfection of the unit, spray the disinfection agent onto a piece of soft cloth / 
cotton and clean the handpiece and power adapter.  
 
Do not use abrasives or solvents as these may damage the plastic parts of the unit! 

Do not spray directly onto handpiece or in button / charge socket! 

 

 

 
 

3.

MAINTENANCE, CHARGING AND REPLACING THE BATTERY

 

 

The charging station can be used as a stand and only when the battery is discharged, 
the charging adapter can be switched on. 
- After polymerization - place the handpiece on the charging base. 
- If the battery is discharged to 5-10%, the display starts flashing, which means the 
battery is almost discharged. If operation continues, the battery is completely 
depleted after a few cycles - 

“Lb” (Low Battery) appears on the display. In this case, 

place the handpiece on the charging station - The CHARGE indicator lights up. While 
charging, the display of the handpiece shows 3 consecutive illuminated lines. When 
the battery is fully charged, a long beep sounds for 30 seconds. The three lines are lit. 
The probe then enters sleep mode. 

Содержание BLUEDENT SMART cordless

Страница 1: ...OPERATING INSTRUCTIONS MANUAL BLUEDENT SMART cordless Dental curing light Ref 200 004 A...

Страница 2: ...OF OPERATION VI ROUTINE CARE AND MAINTENANCE VII WARRANTY CONDITIONS VIII SERVICE DATA IX SYMBOLS X FAQ XI DECLARATION OF CONFORMITY XII BLUEDENT DATA I UNIT DESCRIPTION AND FUNCTION Dental curing li...

Страница 3: ...ing time 10 20 30sec 1 9 min 5 7 Audible signal every 10 sec 1 min according to the selected mode Possibility to turn on off the sound signal 8 Working ability after fully charged battery up to 4000se...

Страница 4: ...connect the device from the mains after completing the procedures Do not use or store the device in a dusty environment Do not expose the device to direct sunlight Do not spray disinfectant directly i...

Страница 5: ...lectronic equipment according to local regulations Proper disposal of equipment that is no longer used prevents negative consequences for the environment and human health In accordance with the requir...

Страница 6: ...eyes and skin with intense light carries a risk of damage from light and heat The light should never be directed at the eyes Irradiation should be limited to the workplace area The special safety gogg...

Страница 7: ...itive drugs persons undergoing cataract surgery persons with retinal diseases people with allergies people who have recently undergone cosmetic surgery on the face or lips including injections of hyal...

Страница 8: ...pproved by the Ministry of Health and sterilized by autoclaving at 134 C for 5 minutes 1 For polymerization of posterior teeth attach as shown 1 Protection cap For polymerization of anterior teeth gen...

Страница 9: ...lymerization P PULSE mode 10 20 30 sec It emits pulsating light which helps reduce shrinkage in the photocomposite It also reduces the patient s thermal sensation 1 Curing mode display 2 Timer display...

Страница 10: ...ing light is switched off at the end of the day by disconnecting the charging adapter from the mains 9 SLEEP function The wireless handpiece has a sleep function which helps reduce energy consumption...

Страница 11: ...l is ON after holding down the MODE button OFF is displayed and the sound is turned off If the sound signal is OFF after holding down the MODE button ON is displayed and the sound is turned on 13 Ther...

Страница 12: ...directly onto handpiece or in button charge socket 3 MAINTENANCE CHARGING AND REPLACING THE BATTERY The charging station can be used as a stand and only when the battery is discharged the charging ad...

Страница 13: ...the probe The installation of a new battery is done in the reverse order The contact surfaces for charging the wireless probe and the charging station should be cleaned periodically with a cotton clo...

Страница 14: ...device the battery must be removed immediately and sent to the manufacturer s service where the damages are established In case of such damage the warranty is lost 3 2 No damages or complaints are acc...

Страница 15: ...fits in case of damage or incompleteness operation of the device whatever the cause 6 In case of disputes in connection with the application and interpretation of this Operating instruction manual the...

Страница 16: ...g life Pay attention to the instructions accompanied by this symbol Fragile Product with CE mark Applied part type BF according to the classification for electrical safety Ohm electric resistance unit...

Страница 17: ...to clean the tip of the fiberoptic lightguide with a non metallic object and to clean the adhered composite or other with alcohol If the result is poor send the device to the service There is no sound...

Страница 18: ...rdless Dental curing light 18 The power cord or AC adapter is damaged Send to the service Liquid entering the device Send to the service After falling to the floor the device does not work properly Se...

Страница 19: ...rt 1 2 General requirements for basic safety and essential performance Collateral Standard Electromagnetic disturbances Requirements and tests EN 62304 2018 Medical device software Software life cycle...

Страница 20: ...SMART cordless Dental curing light 20 XII BLUEDENT DATA SN LOT DATE OF PRODUCTION QC DATE OF PURCHASE Last revision 01 09 2021 Follow www bglight com for updated revision of this Operating instruction...

Отзывы: