MAINTAINENCE: CLEANING / DISINFECTING / STERILIZATION

(Your Slimax unit is maintenance-free. No periodic maintenance is required)

Manual Cleaning and Disinfection of the Light Guide

Place the light guide for the specified application time into the solution, making sure that it is 

completely covered (as needed, using ultrasonic support or careful brushing with a soft brush). 

A neutral enzymatic cleaning agent is recommended.

Remove the light guide from the solution and rinse thoroughly (at least 10 seconds) in water with 

low germ count.

To disinfect; place the cleaned light guide for the specified application time into the solution, mak-

ing sure that it is completely covered. 

Please note: 

Disinfectants containing o-phthalaldehyde are 

recommended.

Remove the light guide from the solution and rinse thoroughly (at least 10 seconds) in water with 

low germ count.

Dry the light guide with a clean soft cloth.

Steam Sterilization

Check the light guide.

Effective cleaning and disinfection are absolutely essential requirements for effective sterilization. 

Only steam sterilization is approved.

Maximum sterilization temperature is 134 °C (273 °F)

Sterilization time (exposure time at sterilization temperature) at least 20 min. at 121°C (250 °F) or 

at least 3 min. at 132 °C (270 °F) /134 °C (273 °F)

Check

Before using the light guide again, check it for damaged surfaces and discoloration;  

do

 

not

 use damaged light guides. 

How to Clean Charger, Main board, and Glare Shield

To disinfect all components, spray the disinfectant on a towel and use it to disinfect the unit. Do 

not spray the disinfectant directly on the Curing Light Controller or the Charger.

Disinfection agents must not enter the unit.

Clean the protecting glass of the Curing Light Controller with a soft and lint-free cloth.

Dry residual disinfectants on the charger, the Curing Light controller and the glare Shield with a 

soft and lint-free cloth.

Solvents or abrasive cleaners 

should

 

not 

be used at any time for cleaning any of the Curing Light 

System components!

Cleaning agents must not enter the Curing Light Controller.

Make sure that the charge contact pins remain dry and are not contacted by metallic or greasy 

parts.

Care must be taken to avoid bending the charge contact pins during drying. Wet charge contact 

pins will cause an operating error (malfunction message: the LED showing the operating status in 

the charger glows red steadily).

Storage of the Slimax during Extended Periods of Non-Use

If the Slimax is not being used for an extended period of time - e. g., during vacation - fully charge 

the battery prior to departure or keep the Curing Light Controller inserted in the operational 

charger. A safety switch within the battery prevents a total discharge.

Discharged or nearly discharged batteries must be recharged as soon as possible.


Содержание SLimax-C

Страница 1: ...LED Curing Light System Operators Manual SLIMAX C SLIMAX CPLUS ...

Страница 2: ...ctly on or towards unprotected gingiva or skin Adjust curing techniques in accordance with the increase in light energy Do not look directly at the light emitted from this LED curing unit Do not use this device without suitable eye protection for the operator assistant and patient Person s having a history of photosensitive reactions or who are using photosensitizing drugs should not be exposed to...

Страница 3: ...onds curing time Users can either select Full Power Ramping Power or Pulse This slim light weight curing light is ergonomically designed for user s comfort and flexibility It s solid aluminum body sets it apart from the mass of plastic competitors The quality durability and ease of applied infection control techniques make this curing unit a state of the art instrument The curing light Slimax C co...

Страница 4: ...re the exposure time is over CURING LIGHT CHARGE BASE Included with the Slimax C Plus is a built in radiometer See Fig ure 2 It is located on the Charge base for easy access to measure the instant output of your curing light To use the radiometer please ensure that the charging base is plugged into a power outlet Turn the curing light on and activate the light Then insert the light guide into A Se...

Страница 5: ...ds LP Regular Output 1400 mW cm2 10 HP Boost Output 2000 mW cm2 10 Select Output 1 Press button 2 continuously 2 Press button 1 to select HP or LP TROUBLESHOOTING Problem Cause s Solution s No indicator lights on the controller Battery Not Charged Battery is improperly connected Recharge the battery Unscrew the battery to clean the circuit contact dot of the battery and the two contact pins locate...

Страница 6: ...the light guide again check it for damaged surfaces and discoloration do not use damaged light guides How to Clean Charger Main board and Glare Shield To disinfect all components spray the disinfectant on a towel and use it to disinfect the unit Do not spray the disinfectant directly on the Curing Light Controller or the Charger Disinfection agents must not enter the unit Clean the protecting glas...

Страница 7: ...of this warranty Beyes liability is limited to at Beyes option to replace the defective product or part thereof or reimbursement of the actual cost of the defective product In order to take advan tage of this limited warranty the defective product must be returned to Beyes Dental Canada Inc In no event shall Beyes be liable for any indirect incidental or consequential damages BeyesCanada warrants ...

Страница 8: ... rang of the charge 100 230V 50 60Hz Power input 10 W 5V 2A Dimensions Diameter 96 mm 3 8 in High 75 mm 2 95 in Weight 6 4 oz 0 4lb Charger Federal law restricts this device to sale by or on the order of a dentist physician or any other practitioner licensed by the law of the states in which he or she practices to use or order the use of this device Printed in Canada I115EN REV2E1903 0197 Beyes De...

Отзывы: