BEYES SLimax-C Скачать руководство пользователя страница 5

5

RECOMMENDED CURING TIME BASE ON THE BOOST POWER

Due to the high light output of the Slimax Curing Light System, the normal exposure times for conventional 

units can be cut in half without compromising polymerization performance.

Beyes Dental Canada TESTING INDICATES THE FOLLOWING RECOMMENDED CURING TIMES:

Universal composite shades A3 and lighter (2-mm depth).....................5 seconds

Universal composite shades A3.5 and darker............................................10 seconds

Bonding agents, universal resin luting cements

and Self-etch, self-adhesive cements....…...................................................5 seconds

Note:

 It is highly recommended that curing capability be tested using a Hardness Disk

to ensure a complete cure.

CURING LIGHT CONTROLLER: 

The efficiency of your new Slimax Curing Light is designed to provide an entire days use when fully charged.  

A fully charged battery will yield 850 ten-second cures. 

Boost and Regular Light Output :

The Slimax Curing Light offers you two tiers of light intensity to meet your daily needs 

LP = Regular Output

      1400 mW/cm2 ± 10%          

HP = Boost Output

         2000 mW/cm2 ± 10% 

Select Output:

1. Press button “2” continuously 

2. Press button “1” to select HP or LP        

TROUBLESHOOTING

Problem

Cause(s)

 

Solution(s)

No indicator lights on the controller

• Battery Not Charged

• Battery is improperly connected

√ 

Recharge the battery

√ 

Unscrew the battery to clean the 

circuit contact dot of the battery and 

the two contact pins located at the end 

of the unit

The battery does not charge

• Check Power Supply and Charger 

Base for proper 

connections

• Check that Controller is 

seated properly in Charger Base

√ 

Ensure there is a good connection 

electrical outlet and plug

 Clean the battery electronic contact 

and charge needle inside of the charger 

base

The controller has not been used for a 

long time and will not turn on

• Battery no charged

 Place Controller in the charger and 

re-charge the battery.

Low light intensity

• The light guide may be not clean

√ 

Clean the light guide and the 

protecting glass in the light guide 

mounting hole

Cannot activate the light by pressing  

the ON/OFF button

• LED light is burned out

√ 

Call Beyes Dental Canada Inc. for an 

authorized service center

Содержание SLimax-C

Страница 1: ...LED Curing Light System Operators Manual SLIMAX C SLIMAX CPLUS ...

Страница 2: ...ctly on or towards unprotected gingiva or skin Adjust curing techniques in accordance with the increase in light energy Do not look directly at the light emitted from this LED curing unit Do not use this device without suitable eye protection for the operator assistant and patient Person s having a history of photosensitive reactions or who are using photosensitizing drugs should not be exposed to...

Страница 3: ...onds curing time Users can either select Full Power Ramping Power or Pulse This slim light weight curing light is ergonomically designed for user s comfort and flexibility It s solid aluminum body sets it apart from the mass of plastic competitors The quality durability and ease of applied infection control techniques make this curing unit a state of the art instrument The curing light Slimax C co...

Страница 4: ...re the exposure time is over CURING LIGHT CHARGE BASE Included with the Slimax C Plus is a built in radiometer See Fig ure 2 It is located on the Charge base for easy access to measure the instant output of your curing light To use the radiometer please ensure that the charging base is plugged into a power outlet Turn the curing light on and activate the light Then insert the light guide into A Se...

Страница 5: ...ds LP Regular Output 1400 mW cm2 10 HP Boost Output 2000 mW cm2 10 Select Output 1 Press button 2 continuously 2 Press button 1 to select HP or LP TROUBLESHOOTING Problem Cause s Solution s No indicator lights on the controller Battery Not Charged Battery is improperly connected Recharge the battery Unscrew the battery to clean the circuit contact dot of the battery and the two contact pins locate...

Страница 6: ...the light guide again check it for damaged surfaces and discoloration do not use damaged light guides How to Clean Charger Main board and Glare Shield To disinfect all components spray the disinfectant on a towel and use it to disinfect the unit Do not spray the disinfectant directly on the Curing Light Controller or the Charger Disinfection agents must not enter the unit Clean the protecting glas...

Страница 7: ...of this warranty Beyes liability is limited to at Beyes option to replace the defective product or part thereof or reimbursement of the actual cost of the defective product In order to take advan tage of this limited warranty the defective product must be returned to Beyes Dental Canada Inc In no event shall Beyes be liable for any indirect incidental or consequential damages BeyesCanada warrants ...

Страница 8: ... rang of the charge 100 230V 50 60Hz Power input 10 W 5V 2A Dimensions Diameter 96 mm 3 8 in High 75 mm 2 95 in Weight 6 4 oz 0 4lb Charger Federal law restricts this device to sale by or on the order of a dentist physician or any other practitioner licensed by the law of the states in which he or she practices to use or order the use of this device Printed in Canada I115EN REV2E1903 0197 Beyes De...

Отзывы: