4. Precaution

4.1 Notice when using equipments

4.1.1 Keep the scaler clean before and after operation.
4.1.2 The handpiece, scaling tip, torque wrench, endo wrench and endochuck must be sterilized 
before each treatment.
4.1.3 Don’t screw or unscrew the scaling tip and endochuck when stepping on the foot switch.

4.1.4 The scaling tip must be fastened and there must be fine spray or dripped from the tip when 

operating.

4.1.5 Change to a new one when the tip and ulrasonic file are damaged or worn excessively.
4.1.6 While scaler working, the heat of scaling tip will become higher if there is no water flowing out. 
Please keep the water flow smoothly.

4.1.7 Don’t twist the tip and endo chuck or rub them.
4.1.8 Don't use impure water source and be sure not use normal brine instead of pure water 
source.
4.1.9 If using the water source without hydraulic pressure, the water surface should be one meter 
higher than the head of patient.
4.1.10 Ensure the connector of handpiece and the socket of the cable are dry before installing the 
handpiece.
4.1.11 Don’t pull the cable forcibly in case of the handpeice falling off the cable.
4.1.12 Don’t’ knock or rub the handpiece.
4.1.13 Please put the power plug into the socket easy to pull out, to make sure it can be pull out in 
emergency.
4.1.14 The device can only be equipped with the special power supply of original manufacturer.
4.1.15 The power supply is NOT waterproof. Please keep it fry and away from the water.
4.1.16 After operating, turn off the power, then pull out the plug.
4.1.17 We are only responsible for the safety on the following conditions:

•  The maintenance, repairment and modification are made by the manufacturer or the the 

authorized dealer.

• 

The changed components are originated of “BEYES” and are operated according to instruction 
manual.

4.1.18 The internal screw thread of the scaling tips produced by some maunactures is coarse, rusty 
and collapsed. This will damage the external screw thread of the handpiece irretrievably. Please 
use “BEYES” brand scaling tips.

4.2 Contraindication

4.2.1 The hemophilia disease patient is not allowed to use this equipment.
4.2.2 The patients or doctors with heart pacemaker are forbidden to use this equipment.
4.2.3 The patient with heart disease, pregnant women and children should be cautious with using 
the equipment.

•  [Note 3] If the scaling tip has been screwed on tightly and there is fine spray too, the following 

phenomena show that the scaling tip is damaged:

 

a. The vibrating intensity and the water atomization degree become weak obviously.

 

b. During treatment, it produces the sound like "buzz" from the scaling tip.

Содержание Comfort-Sonic P6 Plus

Страница 1: ...Comfort Sonic Ultrasonic Scaler P6 Plus Instruction for Use...

Страница 2: ...3 2 Sterilization of scaling tips and endochuck P10 3 3 Serilization of torque wrench and edo wrench P10 3 4 Cleaning of tips endochuck torque wrench and endo wrench P10 3 5 Sterilization and cleanin...

Страница 3: ...for scaling These features make COMFORT SONIC become a new generation product in the world dental market 1 2 Components 1 2 1 The components of the machine are listed in the packing list The scaling...

Страница 4: ...gree against water used on the foot switch IPX1 Degree of safety of application in the presence of a Flammable Anesthetic Mixture with air or with Oxygen or Nitrous Oxide Equipment is not suitable for...

Страница 5: ...icture 2 picture 2 picture 1 b Socket for foot switch Socket for power supply Water entrance Power switch Water control knob Plug of foot switch Foot switch socket Power supply Power cord plug Power c...

Страница 6: ...chable handpiece are showed in picture 4 e Sketch map for how to install tip and endpchuck with wrench are showed in picture 5 LED lamp Nipple Light pipe Handpiece decorative ring The connector of cab...

Страница 7: ...ugh to the power See picture 2 Switch on the main unit then the scaling indicator and the leading five lights of power regulator shine Step on the foot switch the tip begins to vibrate and the LED lam...

Страница 8: ...the high temperature and pressure The connector of the cable Connect the handpiece with the water source and power supply of the main unit LED lamp Light pipe Clean them with purified water and steril...

Страница 9: ...responsible for any damage of the torque wrench directly or indirectly made by any way in the above items 3 Sterilization and maintenance 3 1 Sterilization of detachable handpiece 3 1 1 Autoclaved in...

Страница 10: ...Clean the water pipe by multi function syringe note 2 3 6 1 Troubleshooting The following sterilizing methods are forbiden a Putting handpiece into any liquid for boiling b Dipping handpiece in disinf...

Страница 11: ...nge to a new O ring The ultrasonic doesn t vibrate The screw is loosen Tighten it Endochuck is damaged Change to a new endochunck LED light doesn t work Poor contact Contact tightly Something wrong wi...

Страница 12: ...he socket easy to pull out to make sure it can be pull out in emergency 4 1 14 The device can only be equipped with the special power supply of original manufacturer 4 1 15 The power supply is NOT wat...

Страница 13: ...don t invert it 4 4 2 Don t put it together with dangerous goods during transportation 4 4 3 Avoid solarization and getting wet in rain and snow during transportation 4 5 Working condition Enironment...

Страница 14: ...h upon examination by Beyes prove to be defective This warranty extends only to the first retail purchaser of a product and is not transferable or assignable Replacement components or products may be...

Страница 15: ...zed to sell the product at issue in the geographic territory where the purchaser is located or is not authorized to sell the product at issue within the medical animal health or dental market as the c...

Страница 16: ...of the states in which he or she practices to use or order the use of this device 0197 Beyes Dental Canada Inc 23 595 Middlefield Road Toronto Ontario M1V 3S2 Canada Tel 1 855 603 1888 Fax 1 855 720 1...

Отзывы: