17
6. Trouble Shooting
Problems
Possible causes
Solutions
No power and no signal on
the screen after pressing
power on
1. The battery is placed
incorrectly
2. The battery has been
depleted
1. Re-install the battery.
2. Recharge the battery.
The length of the root canal
cannot be measured.
1. The connection of the
measuring wire is loose
2. The measuring wire is
broken
Confirm the measuring wire is
plugged firmly, link the lip hook
with the file clip to check if the
measuring wire is broken.
No sound of alarm.
Volume is set at “mute”
Adjust the sound level.
The charging LED indicator
goes out.
1.The adapter is not
connected properly
2. A faulty adapter with
excessive output is being
used.
3. The battery is not
installed well.
4. The battery has been
damaged.
Reconnect the adapter.
2.Change the adapter. Only
the original adapter should be
used.
3.Reinsert the battery and
then reconnect the adapter.
4.Change the battery and then
reconnect the adapter.