![Beurer IPL 8500 Скачать руководство пользователя страница 30](http://html.mh-extra.com/html/beurer/ipl-8500/ipl-8500_instructions-for-use-manual_2744210030.webp)
30
Table 4
IEC 60601-1-2: 2014 4
th
Edition
Recommended separation distances between portable and mobile RF communications equipment and the device. The device is intended for use
in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the device can help prevent elec-
tromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the
device as recommended below, according to the maximum output power of the communications equipment.
Rated Maximum Output
Power of Transmitter
(W)
Separation distance according to frequency of transmitter (m)
0.15 MHz to 80 MHz
d
= 3.5/3
=1.17
80 MHz to 800 MHz
d
= 3.5/6
=0.25
80 MHz to 800 MHz
d
= 3.5/10
=0.35
800 MHz to 2,7 GHz
d
= 7/10
=0.70
0.01
0.12
0.058
0.035
0.07
0.1
0.37
0.18
0.11
0.22
1
1.17
0.58
0.35
0.7
10
3.70
1.83
1.11
2.21
100
11.67
5.8
3.5
7
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d
in meters (m) can be estimated
using the equation applicable to the frequency of the transmitter, where
P
is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer.
NOTE 1
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures,
objects and people.