Beurer IH 58 Скачать руководство пользователя страница 6

6

• Connect the other end of the compressed air hose 

[6] to the nebuliser hose connection [5] by turning it 

slightly.

Switching on the nebuliser

To turn on the nebuliser, proceed as follows:

• Plug the micro USB cable included in delivery into the 

micro USB port on the nebuliser. Plug the other end of 

the micro USB cable into the USB port on the mains 

adapter and insert the mains adapter into the socket.

• Press the ON/OFF button on the nebuliser. The ON/

OFF button illuminates blue. The nebuliser is now op-

erational. 

6. Operation

 Important

• For hygiene reasons, it is essential to clean the atom-

iser [7] and the accessories after each treatment and 

to disinfect them after the last treatment of the day.

•  The accessories may only be used by one person; use 

by several people is not recommended.

• If the treatment involves inhaling several different 

medications one after the other, please be aware that 

the atomiser [7] must be rinsed under warm tap water 

following every usage. See “Cleaning and disinfection” 

on page 8.

• Please observe the notes on changing the filter in 

these instructions for use.

• Check that hose connectors are firmly attached to the 

nebuliser [5] and the atomiser [7] before each use of 

the device.

• Before use, check the device is working correctly by 

briefly switching on the nebuliser (together with the 

connected atomiser, but without medication). If air 

comes out of the atomiser [7], the device is working.

6.1 Inserting the atomiser insert

• Open the atomiser [7] by twisting the top anticlock-

wise against the medication container [8]. Place the 

atomiser insert [9] into the medication container [8].

•  Ensure that the cone for administering medication fits 

well on the cone for the air duct inside the atomiser.

6.2 Filling the atomiser 

• Fill with an isotonic saline solution 

or pour the medication directly into 

the medication container [8]. Avoid 

overfilling. The maximum recom-

mended filling quantity is 6 ml.

• Use medication only as instructed 

by your doctor and ask about the 

appropriate inhalation period and 

quantity for you.

• If the prescribed quantity of medi-

cation is less than 2 ml, top this up 

to at least 2 ml with isotonic saline solution. Dilution 

is also necessary with viscous medications. Here too, 

please observe the instructions of your doctor.

6.3 Closing the atomiser

• Close the atomiser [7] by twisting the top clockwise 

against the medication container [8]. Ensure that the 

connection is correct. 

6.4 Connecting accessories to the atomiser

• Connect the atomiser [7] to the desired accessory 

(mouthpiece [10], adult mask [11], children’s mask [12] 

or nosepiece [13]).

In addition to the accompanying adult [11] and children’s 

mask (from 2 years of age) [12], you also have the option 

of using baby mask (from ages 0-3 years) with your ne-

buliser. The baby mask is not included in delivery with the 

IH 58. You can find an overview of all replacement items 

in chapter “12. Replacement parts and wearing parts”.

 Note

The most effective form of nebulisation is by using the 

mouthpiece. Nebulisation using a mask is only recom-

mended if it is not possible to use a mouthpiece (e.g. 

for children who are not yet able to inhale medication 

using a mouthpiece).

8

9

Содержание IH 58

Страница 1: ...IH 58 EN Nebuliser Instructions for use...

Страница 2: ...recovery Ask your doctor or pharmacist for fur ther information about the potential applications The device is suitable for domestic use Inhaler medica tion should only be used following instructions...

Страница 3: ...e dosage and the frequency and duration of inhalation Only use medication prescribed or recommended by your doctor or pharmacist Please note For treatment only use parts indicated by your doctor accor...

Страница 4: ...We therefore recommend that you obtain a replacement device or medication the latter should be agreed with your doctor Should you require an adapter or extension lead this must meet the applicable saf...

Страница 5: ...p with filter 5 Hose connector Overview of atomiser and accessories 6 Compressed air hose 7 Atomiser 8 Medication container 9 Atomiser insert 10 Mouthpiece 11 Adult mask 12 Children s mask 13 Nosepiec...

Страница 6: ...twisting the top anticlock wise against the medication container 8 Place the atomiser insert 9 into the medication container 8 Ensure that the cone for administering medication fits well on the cone f...

Страница 7: ...sensitive bronchial system medications containing essential oils may under certain conditions cause an acute bronchos pasm a sudden cramp like restriction of the bronchi with shortness of breath Consu...

Страница 8: ...container 8 Reassembly is carried out in reverse order Cleaning Important The device must be switched off disconnected from the mains and allowed to cool down each time before cleaning The atomiser an...

Страница 9: ...rs and their accesso ries are affected by a certain amount of wear and tear when used and hygienically prepared on a frequent ba sis Over time this can lead to a change in the aerosol which can have a...

Страница 10: ...hould not be left alone Inhalation us ing the mask takes longer This is due to technical reasons You breathe less medication per breath through the holes of the mask than via the mouthpiece The aeroso...

Страница 11: ...guaranteed We reserve the right to make technical changes to improve and develop the product This device and its accesso ries comply with the European standards EN60601 1 EN60601 1 2 CISPR 11 IEC61000...

Страница 12: ...r local retailer in the first in stance see the attached International Service list of service addresses The buyer will then receive further information about the processing of the warranty claim e g...

Отзывы: