Problem/

question

Possible cause/remedy 

The atomiser 

produces no 

or too little 

aerosol.

4.  Unsuitable medication fluid added 

for nebulisation (e.g. too viscous). 

The medication fluid should be pre-

scribed by the doctor.

The output is 

too low.

Kinked hose, clogged filter, too much 

inhalation solution.

What med-

ications are 

suitable for 

inhaling?

Please consult your doctor in rela-

tion to this matter.

 

As a rule, all medication that is 

suitable and approved for device 

inhalation can be inhaled.

There is inha-

lation solution 

residue in the 

atomiser.

This is normal and is due to technical 

reasons. Stop inhalation once the 

atomiser starts to make a notably 

different sound.

What should 

be taken into 

account when 

using the 

device with 

infants and 

children?

1.  On infants and children, the mask 

should cover the mouth and nose 

to ensure effective inhalation. 

2.  On children, the mask should also 

cover both the nose and mouth. 

It is not a good idea to carry out 

nebulisation on someone who 

is sleeping, as in this case not 

enough of the medication will 

reach the lungs.

Note:

 Inhalation should only be car-

ried out under the supervision of an 

adult and with their assistance and 

the child should not be left alone.

Inhalation us-

ing the mask 

takes longer.

This is due to technical reasons. You 

breathe less medication per breath 

through the holes of the mask than 

via the mouthpiece. The aerosol is 

mixed with ambient air via the holes.

Problem/

question

Possible cause/remedy 

Why should 

the atomiser 

be replaced 

regularly?

There are two reasons for this:

1.  To ensure a therapeutically effec-

tive particle spectrum, the nozzle 

hole must not exceed a specific 

diameter. Due to mechanical and 

thermal stresses, the plastic is 

subject to a certain degree of 

wear. The atomiser insert [9] is 

particularly sensitive. This can also 

alter the composition of the aero-

sol droplets, which directly affects 

the effectiveness of the treatment.

2.  Regularly changing the atomiser 

is also recommended for hygiene 

reasons.

Should each 

person have 

their own 

atomiser?

Yes, this is absolutely essential for 

hygiene reasons.

11. Technical specifications

Type

IH 58

Dimensions  

(W x H x D)

110 x 62 x 47 mm

Weight

176 g

Operating 

pressure

Approx. 0.25 to 0.5 bar

Atomiser filling 

volume

Min. 2 ml

Max. 6 ml

Medication 

flow rate

Approx. 0.25 ml/min.

Sound pressure

Max. 45 dBA 

(acc. to DIN EN 13544-1 section 26)

Mains connection Input: 100 – 240 V~; 50 – 60 Hz; 0.5 A

Output: 5 V 

; 2 A

Expected 

service life

400 h

Operating 

conditions

Temperature: +10 °C to +40 °C

Relative humidity: 10% to 95%

Ambient pressure: 700 to 1060 hPa

Содержание IH 58

Страница 1: ...IH 58 EN Nebuliser Instructions for use...

Страница 2: ...recovery Ask your doctor or pharmacist for fur ther information about the potential applications The device is suitable for domestic use Inhaler medica tion should only be used following instructions...

Страница 3: ...e dosage and the frequency and duration of inhalation Only use medication prescribed or recommended by your doctor or pharmacist Please note For treatment only use parts indicated by your doctor accor...

Страница 4: ...We therefore recommend that you obtain a replacement device or medication the latter should be agreed with your doctor Should you require an adapter or extension lead this must meet the applicable saf...

Страница 5: ...p with filter 5 Hose connector Overview of atomiser and accessories 6 Compressed air hose 7 Atomiser 8 Medication container 9 Atomiser insert 10 Mouthpiece 11 Adult mask 12 Children s mask 13 Nosepiec...

Страница 6: ...twisting the top anticlock wise against the medication container 8 Place the atomiser insert 9 into the medication container 8 Ensure that the cone for administering medication fits well on the cone f...

Страница 7: ...sensitive bronchial system medications containing essential oils may under certain conditions cause an acute bronchos pasm a sudden cramp like restriction of the bronchi with shortness of breath Consu...

Страница 8: ...container 8 Reassembly is carried out in reverse order Cleaning Important The device must be switched off disconnected from the mains and allowed to cool down each time before cleaning The atomiser an...

Страница 9: ...rs and their accesso ries are affected by a certain amount of wear and tear when used and hygienically prepared on a frequent ba sis Over time this can lead to a change in the aerosol which can have a...

Страница 10: ...hould not be left alone Inhalation us ing the mask takes longer This is due to technical reasons You breathe less medication per breath through the holes of the mask than via the mouthpiece The aeroso...

Страница 11: ...guaranteed We reserve the right to make technical changes to improve and develop the product This device and its accesso ries comply with the European standards EN60601 1 EN60601 1 2 CISPR 11 IEC61000...

Страница 12: ...r local retailer in the first in stance see the attached International Service list of service addresses The buyer will then receive further information about the processing of the warranty claim e g...

Отзывы: