3. Closing the atomiser

• To close the atomiser [7], twist the upper section 

clockwise against the medicine container [9]. Ensure 

the parts are connected correctly.

4. Connecting the accessories to the atomiser

• Connect the required accessory part (mouthpiece [10], 

adult mask [11] or child mask [12]) to the atomiser [9].

In addition to the accompanying adult [11] and 

children’s mask (from 2 years of age) [12], you also 

have the option of using baby mask (from ages 0-3 

years) with your nebuliser. The baby mask is not inclu-

ded in delivery with the IH 18. You can find an overview 

of all replacement items in chapter "12. Replacement 

parts and wearing parts".

 Note

The most effective form of nebulisation is by using the 

mouthpiece. Nebulisation using a mask is only recom-

mended if it is not possible to use a mouthpiece (e.g. 

for children who are not yet able to inhale medication 

using a mouthpiece).

When using a mask to inhale, take care to ensure the 

mask fits well and the eyes are unobstructed.

• Before commencing treatment, pull the atomiser 

upwards out of its holder [7].

• Start the nebuliser using the on/off switch [3].

• A flow of spray from the atomiser indicates that the 

device is functioning properly.

5. Treatment

• When inhaling, sit upright and relaxed at a table and 

not in an armchair to avoid compressing the airways 

and therefore impairing the effectiveness of the treat-

ment.

• Inhale the nebulised medicine deeply.

 Important

The device is not suitable for continuous operation; 

after 20 minutes of operation it must be switched off 

for 40 minutes.

 Note

During treatment, hold the atomiser straight (vertical), 

otherwise the atomisation does not work and correct 

function cannot be guaranteed.

 Important

Essential oils, cough medicines, solutions designed 

for gargling, and drops for application to the skin or 

for use in steam baths are wholly unsuitable for inha-

lation using a nebuliser. These substances are often 

highly viscous and can impair the correct function of 

the device and hence affect the effectiveness of its 

application in the long term.

In the case of oversensitivity of the bronchial system, 

medicines containing essential oils can sometimes 

cause acute bronchiospasm (a sudden, cramp-like 

tightening of the bronchi accompanied by breathless-

ness). Always ask your doctor or pharmacist for advice!

6. End of inhalation

If the spray is only output sporadically or if the noise 

changes during inhalation, you can end the treatment.

•  Switch off the nebuliser after treatment using the on/

off switch [3] and disconnect it from the mains.

 After use, replac

e the atomiser [9] in its holder [7].

7. Clean the device

For information about this, see “Cleaning and disinfec-

tion” on page 7.

7.Changing the filter

In normal operating conditions, the air filter must be 

replaced after approx. 500 operating hours or one year. 

Please check the air filter regularly (after 10-12 nebu-

lisation procedures).

If the filter is very contaminated or blocked, replace the 

used filter. If the filter has become wet, it must also be 

replaced with a new one.

 Caution

• Do not try to clean and re-use the used filter!

•  Only use original filters supplied by the manufacturer. 

Otherwise, your nebuliser may become damaged or 

effective therapy can no longer be guaranteed.

• Do not repair or maintain the air filter while it is in use.

• Never operate the device without a filter.

To replace the filter, proceed as follows:

 Caution

• First switch off the unit and unplug it.

• Allow the unit to cool down.

1. 

Remove filter cap [2] upwards.

 Note

If the filter remains in the unit when the cap is re-

moved, remove the filter from the unit using twee-

zers or similar.

2. Replace the filter cap [2] containing the new filter.

3. Check that the filter cap is firmly in place.

Содержание IH 18

Страница 1: ...IH 18 EN Nebuliser Instructions for use ...

Страница 2: ...chi deep into the lungs Exhale normally Once prepared the device is suitable for reuse The prep aration includes the replacement of all accessory parts including atomiser and air filter as well as disinfection of the surfaces of the device with a standard disinfectant Please note that all accessories should be replaced if the device is used by more than one person We recommend that you replace the...

Страница 3: ...ldren Do not use attachments that are not recommended by the manufacturer Connect the unit only to the mains voltage listed on the type plate Do not dip the unit into water and do not use it in wet rooms Do not allow any liquid to penetrate the unit Protect the unit from strong impacts Never touch the mains cable with wet hands this can cause an electric shock Do not pull the mains plug from the s...

Страница 4: ...gainst dust dirt and moisture Never cover the unit during operation Do not use the unit in very dusty environments Switch the unit off immediately if it is defective or malfunctioning The manufacturer shall not be held liable for damage or injuries caused by improper or incorrect use Repairs Note Never open or attempt to repair the unit yourself as otherwise proper function is no longer guaranteed...

Страница 5: ...t medicines are to be inhaled in suc cession during therapeutic use the atomiser 7 must be rinsed under warm tap water after each use For information about this see Cleaning and disinfec tion on page 7 Follow the instructions for changing the filter con tained in these instructions Before each use check that the hose connections on the nebuliser 4 and the atomiser 7 are tight Before use check that...

Страница 6: ...g and drops for application to the skin or for use in steam baths are wholly unsuitable for inha lation using a nebuliser These substances are often highly viscous and can impair the correct function of the device and hence affect the effectiveness of its application in the long term In the case of oversensitivity of the bronchial system medicines containing essential oils can sometimes cause acut...

Страница 7: ...ver use any substances that may be poisonous when in contact with the skin or mucous membranes or when swallowed or inhaled Use a soft dry cloth and non abrasive cleaning prod ucts to clean the device Do not use abrasive detergents and never immerse the unit in water Important Ensure that no water penetrates inside the unit Do not put the device or accessories in a dishwasher Do not touch the unit...

Страница 8: ...ocal authorities 10 Troubleshooting Problem Ques tion Possible Cause Remedy The atomiser produces little or no aerosol 1 Too much or too little medicine in the atomiser Minimum 2 ml Maximum 8 ml 2 Check that the nozzles are not blocked If necessary clean the nozzle e g by rinsing Use the atomiser again CAUTION Only poke imple ments through the fine holes from the underside of the nozzle 3 The atom...

Страница 9: ...060 hPa Aerosol Properties Flow 5 31 l min Aerosol delivery 0 171 ml Aerosol delivery rate 0 082 ml min Particle size MMAD 3 16 µm The serial number is located on the device or in the battery compartment Subject to technical modifications Particle size diagram Measurements were performed using a sodium fluoride solution with a Next Generation Impactor NGI This diagram may not be applicable for sus...

Страница 10: ...er sonal purposes in the context of domestic use German law shall apply During the warranty period should this product prove to be incomplete or defective in functionality in accord ance with the following provisions Beurer shall carry out a repair or a replacement delivery free of charge in accordance with these warranty conditions If the buyer wishes to make a warranty claim they should approach...

Страница 11: ...11 ...

Страница 12: ...12 601 16_IH18_2019 12 16_05_IM2_BEU_EN Beurer GmbH Söflinger Straße 218 89077 Ulm Germany www beurer com www beurer healthguide com ...

Отзывы: