Recommended separation distances between portable or mobile radio-communication tools
and the BY60 Breast Pump
The BY60 Breast Pump is designed to be operated in electromagnetic environments where RF irradiated disturbances are
kept under control. The client or user of the BY60 Breast Pump can favour electromagnetic interferences prevention by
keeping a minimum distance between portable or mobile RF communication systems (transmitters) and the BY60 Breast
Pump as recommended, according to the maximum output power of radio-communication tools.
Rated maximum
output
power of transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
1.17
P
80 MHz to 800 MHz
4
P
800 MHz to 2,5 GHz
2.33
P
0,01
0.12
0.4
0.23
0,1
0.37
1.26
0.74
1
1.17
4.00
2.33
10
3.7
12.65
7.37
100
11.70
40.00
23.3
For those transmitters whose maximum rated output power is not indicated in the above list, the recommended separation
distance in metres (m) can be calculated by the equation applicable to the transmitter frequency, where P is the maximum
rated output of the transmitter in Watt (W) as declared by the manufacturer.
NOTE 1: For 80MHz and 800MHz, the highest frequency interval is applied.
NOTE 2: These guidelines cannot always be applied. Electromagnetic propagation is altered by absorption and refl ection
of facilities, objects and people.