ELECTROMAGNETIC COMPATIBILITY INFORMATION
Guidance and manufacturer’s declaration - electromagnetic emissions
The BY 60 Breast Pump is designed to be operated in the electromagnetic environment described below. The client or
user of the BY 60 Breast Pump must ensure that the product is used in such environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The BY 60 Breast Pump uses RF energy for its internal operation only.
Consequently its RF emissions are very low and cause no interference of
nearby electronic equipment.
RF emissions
CISPR 11
Class B
The BY 60 Breast Pump is suitable for use in all types of facilities, dome-
stic buildings included, and in those buildings directly connected to the
low voltage public electricity supply system of domestic buildings always
keeping in mind what follows:
Caution: This apparatus / system has to be used by health care professio-
nals only. This apparatus / system could cause radio interferences or alter
operation of nearby equipment. Attenuation measures could be neces-
sary, such as re-orienting or re-positioning of the BY 60 Breast Pump or
shielding of the current position.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complying
Guidance and manufacturer’s declaration - electromagnetic immunity
The BY 60 Breast Pump is designed to be operated in the electromagnetic environment described below. The client or
user of the BY 60 Breast Pump must ensure that the product is used in such environment.
Immunity test
Electrostatic discharges
IEC 61000-4-2
Test level
± 6kV in contact
± 8kV in air
Conformity level
l± 6kV in contact
± 8kV in air
Electromagnetic environment -
guidelines
Floors have to be made in wood, ce-
ment, or ceramic. If fl oors are coated
with synthetic materials relative humi-
dity must range at least at 30%.
Transients / Fast
electric trains
IEC 61000-4-4
± 2kV for
power supply lines
± 1kV for
input/output lines
± 2kV for
power supply lines
± 1kV for
input/output lines
The quality of the mains voltage should
be that of a typical commercial or hos-
pital environment.
Pulses
IEC 61000-4-5
± 1kV in differential way
± 2kV in common way
± 1kV in differential way
± 2kV in common way
The quality of the mains voltage should
be that of a typical commercial or hos-
pital environment.
Voltage holes, voltage
short interruptions and
variations on the power
supply input lines
IEC 61000-4-11
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
<5% U
T
(>95% dip in U
T
)
for 5 sec
0% UT
for 10 ms
40% UT
for 100 ms
70% UT
for 500 ms
0% UT for 5 sec
The quality of the mains voltage
should be that of a typical com-
mercial or hospital environment. If
the user of the BY60 Breast Pump
requires continuous operations even
in case of power supply cutoff, it is
recommended to plug the BY60 Bre-
ast Pump to a uninterruptible power
supply or batteries.
Mains frequency ma-
gnetic field (50/60 Hz)
IEC 61000-4-8
3 A/m
Non applicable
Mains frequency magnetic fi eld
has to be at a level typical of a
commercial or hospital environment.
NOTE
U
T
is the a.c. mains voltage before to application of the test level.