27
• Press the
-
or
+
function button. Select the desired user
memory ( ... ) by pressing the
-/+
function button.
Confirm your selection with the START/STOP button .
• If you press the memory button
M
, the average value of
all the stored measured values in the user memory will be
displayed. If you press the memory button
M
again, the
average value of the morning measurements for the last 7
days will be displayed. (morning: 5.00 – 9.00, display ).
If you press the memory button
M
again, the average
value of the evening measurements for the last 7 days will
be displayed. (evening: 18.00 – 20.00, display ). If you
continue to press the memory button
M
, the most recent
individual measured values are displayed in turn with the
date and time.
• You can delete the memory by pressing and holding the
memory button
M
for 3 seconds.
• To switch off, press the START/STOP button .
• If you forget to switch off the unit, it will switch off auto-
matically after 1 minute.
8. Error message/trouble shooting
In the event of errors, the error message _ appears on the
display.
Error messages may appear if:
• systolic or diastolic pressure could not be measured (
or
appears on the display)
• systolic or diastolic pressure was outside the measure-
ment range ( or
Lo
appears on the display)
• the cuff is fastened too tightly or loosely (
or
appears on the display)
• the pump pressure is higher than 300 mmHg (
appears on the display)
• pumping up takes longer than 160 seconds (
appears
on the display)
• there is a system or device error (
,
,
or
appears on the display)
• the batteries are almost empty
.
In the above cases, you must repeat the measurement.
Make sure that the cuff tube is properly inserted and that
you do not move or talk.
Re-insert the batteries if necessary, or else replace them.
Technical alarm – description
Should the recorded blood pressure (systolic or diastolic) lie
outside the limits specified in the section “Technical speci-
fications”, the technical alarm will appear on the display
indicating either “ ” or “
Lo
”. In such cases, you should seek
medical assistance and check the accuracy of your proce-
dure.
The limit values for the technical alarm are factory set and
cannot be adjusted or deactivated. These alarm limit val-
Содержание BM 47
Страница 90: ...90 4 1 5 B LR6 Beurer 1 1 90 2 91 3 96 4 97 5 98 6 100 7 101 8 102 9 103 10 103 11 105 12 105...
Страница 91: ...91 2 BF EC WEEE Waste Elec trical and Electronic Equipment 21 PAP Storage Transport...
Страница 92: ...92 Operating 0483 CE 93 42 EEC 5 5 30...
Страница 93: ...93 1...
Страница 94: ...94...
Страница 95: ...95 WEEE Waste Electrical and Electronic Equipment...
Страница 96: ...96 3 2 1 3 8 7 6 5 4 9 1 2 3 4 5 6 7 8 9 M 1 10 9 8 5 6 7 2 3 4 1 2 3 4 5 6 7 8 9 10...
Страница 97: ...97 4 1 5 AA LR6 Pb Cd Hg 24 3 M M 4 x AA LR6 1 5 V...
Страница 98: ...98 M 5 2 3 OK 22 35...
Страница 99: ...99 162 795 30 42 5 4 30 0...
Страница 100: ...100 _ 1 5 6 5...
Страница 102: ...102 M M 7 5 00 9 00 M 7 18 00 20 00 M M 3 1 8 _ Lo 300 160 Lo...
Страница 103: ...103 IEC 60601 1 8 8 9 10 BM 47 0 300 60 260 40 199 40 180 3 3 5 8 8 4 30 134 x 102 x 52 5 280...
Страница 105: ...105 11 22 35 162 139 30 42 162 795 12...