40
Beurer BGL60 mg/dL
5
"CTL" appears in the display.
6
Shake the control solution well before use. Unscrew the cap and squeeze out
a drop of solution. Wipe away the first drop and squeeze out an additional
drop.
7
To prevent the remaining control solution in the bottle from becoming con-
taminated through the tip of the bottle in contact with the test strip, do not ap-
ply the drop directly to the test strip. Apply the drop to a clean substrate. Then
put the drop on the slit for receiving blood on the test strip. The solution is
sucked into the slit. Wipe the tip of the bottle using a clean, dry paper tissue.
8
Once the slit is filled with solution, the unit begins the measurement. The unit
counts down for approx. six seconds. Then the measurement result is shown
in the display.
9
Check whether the result is within the specified results range for the control
solution. This results range is printed on the container with the test strips.