39
Note
•
Store the control solution tightly closed at room temperature below 30°C. Do
not refrigerate.
•
Store the control solution out of reach of young children.
•
Do not store the control solution in the medicine cabinet. It may be mistaken
for medicine to be consumed.
Preparations
•
Do not use the control solution once the expiry date has passed. Using control
solution that is out-of-date may lead to incorrect function test results. The ex-
piry date is located on the bottle next to the hourglass symbol
.
•
The control solution can be kept for three months once the bottle is open.
Make a note of when this time will run out (date of o 3 months) on the
label
. The shelf life decreases if the expiry date is reached before the end
of the 3 months (see the date next to the hourglass symbol
).
•
Do not use the control solution if either of these dates has expired (
/
).
•
Allow the measuring device, test strips and control solution to reach room
temperature (+20°C to +25°C).
Performing a function test with control solution
1
Hold the measuring device so that the display is facing you.
2
Insert a test strip, contacts first, into the slot on the measuring device. Ensure
that the front side of the test strip is facing you (see "Test strips" page 19).
3
The unit switches on automatically and briefly shows the start display. As soon
as the flashing blood droplet symbol is displayed, the unit is ready to perform
measurements.
4
Press the memory button [M] to switch to control mode. In control mode, the
measured value is not stored and hence does not distort your measurement
statistics.