88
Table 2
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacturer’s declaration – electromagnetic immunity
The BC 31 is intended for use in the electromagnetic environment specifi ed below.
The customer or the user of the BC 31 should assure that it is used in such an environment.
IMMUNITY test
IEC 60601 test level Compliance level Electromagnetic environment – guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or ceramic tile. If
fl oors are covered with synthetic material, the relative
humidity should be at least 30 %.
Power frequency (50/60 Hz)
magnetic fi eld IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fi elds should be at levels
characteristic of a typical location in a typical commer-
cial or hospital environment.
NOTE:
U
T
is the a.c. mains voltage prior to application of the test level.
Содержание BC 31
Страница 64: ...64 Beurer 1 2...
Страница 66: ...66 1...
Страница 67: ...67 2002 96 EC WEEE Waste Electrical and Electronic Equipment...
Страница 69: ...69 24 M M M 1 12 5 1 5...
Страница 70: ...70 _ 5...
Страница 71: ...71 6 3 180 110 2 160 179 100 109 1 140 159 90 99 130 139 85 89 120 129 80 84 120 80 1999...
Страница 72: ...72 60 M M M 3 7 _ 300 160 IEC 60601 1 8...
Страница 74: ...74 180 2x 1 5 AAA IPX0 AP APG BF EN60601 1 2 93 42 EC EN1060 1 1 EN1060 3 3 IEC80601 2 30 2 30 10 36...
Страница 75: ...75 DE 93 01352 05 06 2012 5 218 89077 OOO 109451 62 2 3 109451 62 2 495 658 54 90 bts service ctdz ru...
Страница 91: ...91...
Страница 92: ...92 750 791 0513 Irrtum und nderungen vorbehalten...