81
Table 2
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacture’s declaration – electromagnetic immunity
The SR BM 1 is intended for use in the electromagnetic environment specified below. The customer or the user of SR BM 1 should assure that it is
used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment – guidance
Electrostatic discharge
(ESD) IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or ceramic tile. If floor
are covered with synthetic material, the relative humi-
dity should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power supply
lines
± 1 kV for input/output
lines
± 2 kV for power supply
lines
± 1 kV for input/output
lines
Mains power quality should be that of a typical com-
mercial or hospital environment.
Surge IEC 61000-4-5
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
Mains power quality should be that of a typical com-
mercial or hospital environment.
Voltage dips, short inter-
ruptions and voltage
variations on power sup-
ply input lines
IEC 61000-4-11
<5 % U
T
(>95 % dip in U
T
)
for 0.5 cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
<5 % U
T
(>95 % dip in U
T
)
for 5 s
<5 % U
T
(>95 % dip in U
T
)
for 0.5 cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
<5 % U
T
(>95 % dip in U
T
)
for 5 s
Mains power quality should be that of a typical com-
mercial or hospital environment. The SR BM 1 can con-
tinue the operation during power mains interruptions
due to the usage of battery.
Power frequency
(50/60 Hz) magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Mains power quality should be that of a typical com-
mercial or hospital environment.
NOTE:
U
T
is the a.c. mains voltage prior to application of the test level.