Electromagnetic Compatibility Information
Table 1
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacture’s declaration – electromagnetic emission
The SR BM 1 is intended for use in the electromagnetic environment specified below. The customer of the user of the SR BM 1 should assure that it is
used in such and environment.
Emission test
Compliance Electromagnetic environment – guidance
RF emissions CISPR 11
Group 1
The SR BM 1 uses RF energy only for its internal function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby electronic equipment.
RF emissions CISPR 11
Class B
The SR BM 1 Blood Pressure Monitor is suitable for use in all establishments, including domestic esta-
blishments and those directly connected to the public low-voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies