MasterCAL
Users Manual
84
Memory Operations
Saving Results
As Found/As Left test results are automatically
saved at the end of each test routine. Any other time
during MEASURE, SOURCE, or MEASURE /
SOURCE you can press the
Save
softkey to save the
data on the display for later review.
After you press
Save
, the calibrator saves the
information on the display and shows a saved result
index number, the date and time, and the percentage
of memory available, as in the following display:
gj49s.eps
If you want to add information to the saved data, the
calibrator has a way for you to do so. If you press the
Continue
softkey, the display prompts you to enter
the instrument tag identifier (
Tag
), instrument serial
number (
S/N
), and operator name (
ID
), as shown in
the following display: