56
Simulating RTDs
Note
Refer to Table 6 for information about RTD
(Resistance-Temperature Detector) types
supported by the calibrator.
Connect the calibrator to the instrument under test
as shown in Figure 20. The figure shows
connections for two, three, or four-wire transmitters.
For three or four-wire transmitters, use the 4-inch
long stackable jumper cables to connect the third
and fourth wires at the source V
Ω
RTD jacks.
Proceed as follows to simulate an RTD (Resistance-
Temperature Detector):
1.
If necessary, press
M
for SOURCE mode.
2.
Press
t
, then select RTD from the menu.
3.
Press the
u
or
d
keys followed by
e
to
select the desired RTD type.
4.
Enter the temperature you want to simulate as
prompted by the display, then press
e
.