Troubleshooting Guide

The following list of encountered problems and solutions will assist you in determining 

what may be causing your patient lift not to function as designed. If you have a problem 

occurring which is not listed below please contact your dealer or technical support for help. 

Do not attempt to repair or replace components or parts on your lift as this may void your 

warranty or cause further problems that may result in patient injury. Stop using your lift 

immediately if it is not functioning correctly or any warning beeps are heard.

I need to re-charge my batteries often or they fail to hold a charge when I charge 

them.

•  Replace the batteries or battery pack as they are at the end of their life cycle

•  Batteries should be changed every 18-24 months depending on usage

The actuator will either go up or down but not both

•  Replace your hand control as it is at the end of its life cycle

•  Your control box may be malfunctioning

There is a grinding sound inside the actuator when lifting

•  Replace your actuator as the internal gears are stripped

The actuator stops and starts while lifting or lowering

•  Replace the batteries or battery pack as they are at the end of their life cycle

My lift will not operate even though it shows the batteries are charged

•  Unplug the AC power cord from the control box as the lift will not work when the 

control box/charger is connected to AC Power.

•  Check to make sure the hand control is properly inserted all the way into the control 

box port.

•  Check to make sure the actuator plug is properly connected and inserted all the way 

into the control box

•  Your hand control needs to be replaced as it is at the end of its life cycle

If there is a problem you could not solve, please contact your dealer for assistance.

Содержание BESTLIFT PL400

Страница 1: ...in this manual All operators should read and understand the instructions for safe and proper operation of the patient lift PL500 model shown https www bestcarellc com warranty registration Scan the QR...

Страница 2: ...o better serve you please record the following information for future use Supplier Name ___________________________________________ Supplier Telephone ________________________________________ Product...

Страница 3: ...ttery pack you will never have to be worried about your lift not being charged and ready for use The PL400 and PL500 are intended to be used with patients who are completely immobile or totally depend...

Страница 4: ...physical contact with the patient in order for the device to carry out its intended Do Not Iron Item may not be smooted or finished with an iron Class II symbol To identify equipment is designed that...

Страница 5: ...ise severely handicapped The patient lift should be used solely for transferring a user patient from one utility beds chairs toilets etc to another The patient lift should not be used for transporting...

Страница 6: ...xposed that may cause bodily harm or injury Prolonged exposure or physical contact with any part of the equipment WILL NOT cause skin irritation allergic reaction etc Using accessories detachable part...

Страница 7: ...0 PACIFIC DRIVE NORCROSS GA 30071 PROPRIETARY AND CONFIDENTIAL THE INFORMATION CONTAINED IN THIS DRAWING IS THE SOLE PROPERTY OF BESTCARE LLC ANY REPRODUCTION IN PART OR AS A WHOLE WITHOUT THE WRITTEN...

Страница 8: ...4 1254 0 49 4 C 699 8 27 6 699 8 27 6 D 643 9 25 4 643 9 25 4 E 998 0 39 3 998 0 39 3 F 887 4 34 9 887 4 34 9 G 570 2 22 4 570 2 22 4 H 270 1 10 6 270 1 10 6 I 545 2 21 5 545 2 21 5 J 271 2 10 7 271...

Страница 9: ...8 mm 26 4 670 8 mm C Base Int Width Opened 39 3 998 mm 39 3 998 mm D Mast Height 51 3 1304 mm 51 3 1304 mm E Front Caster Height 4 7 119 4 mm 4 7 119 4 mm F Lifting Height Lowest 21 9 556 0 mm 21 9 55...

Страница 10: ...mast and align the side holes on the mast with the holes on the bracket Secure the mast to the bracket using the side bolts and cap lock nuts Step 3 Insert the carriage bolt to the front hole of the...

Страница 11: ...over and plastic washers are still attached to the boom Step 4 Insert the round plastic cover and the large plastic washer to the boom as follows Step 6 Secure the boom to the mast by reinstalling the...

Страница 12: ...ION OF BESTCARE LLC IS PROHIBITED 9 8 2017 1 of 1 LOUIS CHOW PIN PL500E Pin 1 1 ALL DIMENSIONS IN MM 1 6 6 30 30 120 120 400 0 1 0 2 0 3 0 5 Step 7 First secure the lower end of the actuator to the br...

Страница 13: ...g Up Power Control Unit 4 3 2 1 5 M5 SCREWS M4 SCREWS UNLESS OTHERWISE SPECIFIED PART NUMBER DATE DRAWN BY 2920 PACIFIC DRIVE NORCROSS GA 30071 9 27 2017 ALL DIMENSIONS IN MM 1 6 6 30 30 120 0 1 0 2 0...

Страница 14: ...e one that serves the needs of the patient while providing the patient with optimal safety Never interfere with the lift unless instructed by the attendant Have a doctor nurse or medical attendant exp...

Страница 15: ...mproperly Keep long charger cord out of walkway where it poses a tripping hazard Have the charger port on the controller facing away from the wall for easy disconnection When removing the battery need...

Страница 16: ...harger must be connected The emergency stop button must be released The battery will not be charged if the emergency button is pressed in LCD Display Panel Signs When the emergency stop button is rele...

Страница 17: ...16 LED Indicator Green In Use Blank Standby Lifting Lowering Operating Lift Using Hand Control Operating Lift Using LCD Display Panel...

Страница 18: ...w the LCD Display Panel will show the sign on the left either when the emergency stop button is released or when a button on the LCD Display Panel is pressed Lift operation is deactivated Battery need...

Страница 19: ...lyester with PU Coating M L XL Liko Guldmann Replacement with Head Support Polyester with PU Coating M L XL Liko Guldmann Disposable Replacement Nonwoven Disposable M L XL Liko Guldmann Disposable Rep...

Страница 20: ...1 Hook up F Point 2 Hook up E Insert F through E 8 Strap Sling A B G H C D E F Point 1 Hook up A C Point 2 Hook up B D Point 3 Hook up E Point 4 Hook up F Point 5 Hook up H Point 6 Hook up G Insert H...

Страница 21: ...he surface 11 Reverse the above procedures when returning the user patient to bed Transfer From Wheelchair to Bed 1 Help the user lean forward slightly then slide the sling down between the chair and...

Страница 22: ...isual inspection prior to use for the additional points below V The periodic inspection interval and the visual inspection points are listed below Note Bestcare assumes that the assembly instructions...

Страница 23: ...propyl alcohol For slings with rigid supports remove the plastic inserts before washing Look for an opening near the plastic inserts for removal Machine wash warm or cold Do not wash with other colors...

Страница 24: ...he user The Emergency Lowering Device is located at the top of the actuator shaft It is intended for use if the actuator or electronics fail to operate while the patient is suspended in midair The dev...

Страница 25: ...g on usage The actuator will either go up or down but not both Replace your hand control as it is at the end of its life cycle Your control box may be malfunctioning There is a grinding sound inside t...

Страница 26: ...suance of a return authorization will also be refused as the authorization will have expired The final disposition of warranty claims will be subject to our inspection of the returned items Bestcare L...

Страница 27: ...NOTES...

Страница 28: ...Bestcare LLC 2920 Pacific Drive Norcross GA 30071 Ver 2021 12 www bestcarellc com 2021 Bestcare LLC...

Отзывы: