Technical Information - Shield LB 8901
64449TI Rev.00, 10/2017
Technical Data
Material housing
316 L
Material shield
Tungsten
Weight with transport cap
approx. 8,0 kg
Weight without transport cap
approx. 6,0 kg
Radiation beam outlet
approx. 11°
Attenuation factor
approx. 170 (Cs-137)
Operating temperature
-40°… +100°C
Maximum permissible activity (Cs-137)
300 mCi / 11100 MBq
NOTICE
The activities indicated here are the maximum activities with which the shield can be loaded
according to the international dose rate limit for the transport of radioactive substances (2000 µSv/h
at the surface of the package and 100 µSv/h at a distance of one meter from the surface). National
dose rate limits for the operation of measurement systems are usually significantly lower and only
allow loading with significantly lower activities. The national provisions must be complied with.
Содержание LB 8901
Страница 1: ...Shield LB 8901 Operating Manual 64449BA2 Rev No 01 10 2018...
Страница 2: ......
Страница 4: ......
Страница 16: ...2 Safety manual LB 8901 Shield 12 64449BA2 Rev 01 10 2018...
Страница 24: ...4 Installation LB 8901 Shield 20 64449BA2 Rev 01 10 2018...
Страница 34: ...5 Maintenance and Repair LB 8901 Shield 30 64449BA2 Rev 01 10 2018...
Страница 41: ...Shield LB 8901 Technical Information 64449TI Rev No 00 10 2017...
Страница 44: ...Technical Information Shield LB 8901 64449TI Rev 00 10 2017 Drawing general view LB 8901 without transport cap...
Страница 45: ...Technical Information Shield LB 8901 64449TI Rev 00 10 2017 Drawing transport cap Drawing flange...
Страница 47: ...Technical Information Shield LB 8901 64449TI Rev 00 10 2017 Signs...