04/03
347
LB
444
7
3.2.2
Shieldings with Pneumatically Operated Lock and
Shutter Switch
NOTE:
Option not available in the USA.
A pneumatic lock with switch contacts indicating the position of the lock is available as a
special version.
The pressurized air moves the shutter to the OPEN position. If the pressurized air is
turned off or in case of failure the moveable shutter is turned back to the CLOSED posi-
tion by the spiral spring.
Pressurized air:
min. 4*10
5
Pa (4 kp/cm
2
)
max. 7*10
5
Pa (7 kp/cm
2
)
Shutter switch:
IP 65
max. 250 V,
40 VA, 1 A
Air quality
clean, as usual for pressurized
air tools, oil free
Temperature range:
-5 to + 60°C
CAUTION-
RADIOACTIVE
MATERIAL
Throttle
valve
Pressurized air
connection
Spring unit
Drive
Shutter
switch
Figure 4: Shielding container type LB 744x with pneumatic locking drive
Do not open spring unit. DANGER!
The pneumatic drive is equipped with a throttle valve. The valve must be set
such that the shielding’s opening and closing process takes at least 2 s; other-
wise the shielding may get damaged!
Содержание LB 444
Страница 1: ...Density Meter LB 444 Id No 32816BA2 Rev No 04 22 04 03 ...
Страница 2: ......
Страница 9: ...04 03 347 LB 444 VII ...
Страница 10: ......
Страница 109: ...04 03 347 LB 444 99 13 4 Mounting Device 90 for Pipe Diameter 21 3 76 1 mm ...
Страница 111: ...04 03 347 LB 444 101 13 6 LB 444 19 module Installation in 19 rack Installation in wall housing ...
Страница 112: ...04 03 347 LB 444 102 ...
Страница 113: ...04 03 347 LB 444 103 ...
Страница 114: ...04 03 347 LB 444 104 ...
Страница 117: ...04 03 347 LB 444 1 14 Ex Certificates for Evaluation Unit ...
Страница 118: ...04 03 347 LB 444 2 ...
Страница 119: ...04 03 347 LB 444 3 ...
Страница 120: ...04 03 347 LB 444 4 15 Ex Certificate for the Detectors ...
Страница 121: ...04 03 347 LB 444 5 ...
Страница 122: ...04 03 347 LB 444 6 ...
Страница 123: ...04 03 347 LB 444 7 ...
Страница 124: ...04 03 347 LB 444 8 ...