Chapter 15 Troubleshooting and correcting faults
Call one or more persons to assist. Otherwise in an emergency
situation the adequate support of the patient is not guaranteed.
The driving tubes and cannulae should be arranged in a bend-free
position. Otherwise in an emergency situation the adequate support of
the patient is not guaranteed.
When operating the manual pump with 1 hand, do not block the valves
with your feet (see valve "2" in Fig. 15-2, page 205).
NOTICE
Seal the connector(s) on the Ikus immediately after removing the
driving tube(s) in order to avoid contaminates from entering the system.
IMPORTANT: In biventricular mode: the blood pumps are driven asynchronously by
the manual pump.
INSTRUCTION
1.
The patient is lying down.
2.
Disconnect the driving tube(s) from the Ikus. To do so, take hold of the release
sleeve and pull this out of the connector.
3.
Connect the driving tube(s) to the manual pump. IMPORTANT: Observe the
colored markings.
4.
Check that the plug is securely connected. To do so, grip the plug body above
the release sleeve and pull on it. (see „4“ in Fig. 15-3, page 206) Do not pull
from the release sleeve, and never from the tube!
5.
Pump steadily and rhythmically at roughly 60 to 80 strokes per minute. Impor
tant: Move the piston so far that the membrane reaches its final position. The
piston need not necessarily be moved to its end position.
6.
Perform a visual check of the blood pump to verify that the membrane is mov
ing and that blood is being pumped.
1
Base plate
2
Valve
3
Connector for driving tube with blue
marking
4
Connector for driving tube with blue
marking
Fig. 15-2
Manual pump
1000721x09 Revision 8
205
Содержание EXCOR Pediatric VAD
Страница 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 43: ...Chapter 3 Summary of Clinical Studies Tab 3 5 Demographic data summary a 1000721x09 Revision 8 33 ...
Страница 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Страница 102: ...Chapter 6 Instructions for use Ikus This page was left blank intentionally 1000721x09 Revision 8 92 ...
Страница 142: ...Chapter 9 Implantation surgical procedure This page was left blank intentionally 1000721x09 Revision 8 132 ...
Страница 162: ...Chapter 12 Anticoagulation therapy This page was left blank intentionally 1000721x09 Revision 8 152 ...
Страница 200: ...Chapter 14 Error Messages and corrective measures This page was left blank intentionally 1000721x09 Revision 8 190 ...
Страница 220: ...Chapter 15 Troubleshooting and correcting faults This page was left blank intentionally 1000721x09 Revision 8 210 ...
Страница 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Страница 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Страница 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Страница 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Страница 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Страница 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Страница 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Страница 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Страница 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...