Chapter 15 Troubleshooting and correcting faults
1
Base
2
Base plate
3
Release sleeve
4
Area above release sleeve
Fig. 15-3
Plug on the driving tube
Fig. 15-4
Examples to operate the manual pump
The manual pump can be operated with both hands or with one hand (placing the
pump between the feet). Alternating between two-handed or one-handed pumping, as
well as using the left or right hand, is allowed. When doing so, care of the patient must
remain ensured.
15.6 Mains failure or breakdown of both control computers
INSTRUCTION
1.
Assess the condition of the patient.
2.
Check the filling and ejection behavior of the blood pump(s).
3.
If possible: ensure support of patient with a replacement Ikus and switch off
the faulty Ikus (see section 15.4: Connecting the patient to a replacement Ikus,
page 202).
If both control computers fail at the same time or if there is a mains failure, the Ikus
cannot generate any specific error messages in the message window. An acoustic
alarm sounds and the indicator lamp on the handle lights up. Depending on the type
of fault, the message
Error: no data from Master.
or
Error: no reaction from Master
appears in the message window.
Ikus is running in emergency pulse mode (see section 15.3: Emergency pulse mode,
page 200).
HOTLINE
Notify Berlin Heart! 866.249.0128
1000721x09 Revision 8
206
Содержание EXCOR Pediatric VAD
Страница 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Страница 43: ...Chapter 3 Summary of Clinical Studies Tab 3 5 Demographic data summary a 1000721x09 Revision 8 33 ...
Страница 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Страница 102: ...Chapter 6 Instructions for use Ikus This page was left blank intentionally 1000721x09 Revision 8 92 ...
Страница 142: ...Chapter 9 Implantation surgical procedure This page was left blank intentionally 1000721x09 Revision 8 132 ...
Страница 162: ...Chapter 12 Anticoagulation therapy This page was left blank intentionally 1000721x09 Revision 8 152 ...
Страница 200: ...Chapter 14 Error Messages and corrective measures This page was left blank intentionally 1000721x09 Revision 8 190 ...
Страница 220: ...Chapter 15 Troubleshooting and correcting faults This page was left blank intentionally 1000721x09 Revision 8 210 ...
Страница 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Страница 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Страница 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Страница 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Страница 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Страница 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Страница 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Страница 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Страница 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...