96
1015002x00
Configuring and Operating the Driving Unit
en
8.3.2
Checking Functionality of a Configured Replacement Driving Unit
8.4
Parameters and Configuration
8.4.1
Default Parameter Settings
The driving unit is set with default values at the factory and whenever driving unit set-
tings are reset.
7.
Check the following items on the driving unit:
•
Did the start test run without errors?
•
Are the displays active?
•
Can an audible alarm be heard when the buttons are pressed?
8.
Check the following items on the panel PC:
•
Is a driving unit – panel PC connection in place?
•
Does the user interface respond when the buttons are pressed?
•
Is the user interface bright and easy to read?
1.
Have the panel PC and driving unit ready.
2.
Connect the panel PC to the mains and switch it on.
3.
Remove the seal plugs.
4.
Connect the driving unit to the mains and insert batteries.
5.
The driving unit will start.
6.
Connect the driving unit and panel PC with the USB cable.
7.
Log into the user interface as EXPERT or MEDICAL.
8.
Check the following items on the driving unit:
•
Are the displays active?
•
Can an audible alarm be heard when the buttons are pressed?
9.
Check the following items on the panel PC:
•
Is a driving unit – panel PC connection in place?
•
Does the user interface respond when the buttons are pressed?
•
Is the user interface bright and easy to read?
10.
Switch off the driving unit. Close the driving tube connectors.
Parameter
Default value
Operating mode
BVAD:
Language
English
Child protection
Off
Tab. 8-1
Default driving unit settings
Содержание EXCOR Adult
Страница 2: ...2 1015002x00 en ...
Страница 12: ...12 1015002x00 Table of Contents en ...
Страница 14: ...14 1015002x00 Contact en ...
Страница 66: ...66 1015002x00 Driving Unit en ...
Страница 94: ...94 1015002x00 User Interface en Fig 7 7 Help window 1 Help display 2 Help button 1 2 ...
Страница 122: ...122 1015002x00 Configuring and Operating the Driving Unit en ...
Страница 126: ...126 1015002x00 Replacing the Driving System en ...
Страница 128: ...128 1015002x00 Weaning and Explantation en ...
Страница 134: ...134 1015002x00 Mobile Use of EXCOR Active en ...
Страница 150: ...150 1015002x00 Baby Buggy Stroller Board en Fig 15 3 Guiding of the driving tubes ...
Страница 154: ...154 1015002x00 Manual Pump en ...
Страница 170: ...170 1015002x00 Identifying Errors and Troubleshooting en ...
Страница 180: ...180 1015002x00 Maintenance Cleaning and Disinfection en ...
Страница 194: ...194 1015002x00 Technical Specifications en ...
Страница 198: ...198 1015002x00 Overview Parameters en ...
Страница 206: ...206 1015002x00 EMC en ...
Страница 208: ...208 1015002x00 Patient ID Labels en ...
Страница 210: ...Abbrevations 210 1015002x00 en ...
Страница 212: ...212 1015002x00 en ...