tive metal
pieces and take out the battery, replace a new one. While changing, be careful to
make sure of the correct connection of electrode of the battery.
4.9.4 About recharging: there are three colors of charging light to indicate the vlotage
of battery, red means the battery needs continous recharging, yellow means the
volt age is 70% full and green means the voltage is 100% full.
5.1 Main directories
Under power on state, firstly press [I/O] to turn it off and then press [I/O] to turn
it on again.When the indicator is in self-checking "9"
=
"0" process, press [I/O]
and then [
0
] in turn (notice: be sure press [I/O] and then [
0
] before
the self-checking is finished ), at this time the indicator will display “F1” .
If it is under power off state, process the operation mentioned above after 30 min
utes warming up.
Press [
] successively, the indicator will select the main directories in F2
F3
=======
F1
=
F2 sequence.
Press [ENTER] to enter any subdirectory under the main directories.
Press [ I/O ] and [C] at the same time to back to the entrance self-checking .
5.2 Subdirectories and selection
When the indicator display "F1"or "F2" or "F3", enter subdirectories by
pressing [ENTER]. make selecation in subdirectories by pressing [
] and
make the indicator withdraw the subdirectories by pressing [ EXIT].
5.2.1 Subdirectories F1
F1.1: weight unit selection, with 4 options
F1.1=1 only for choosing "kg" unit
F1.1=2 "kg" and "lb" unit choosing and converting, whenever turn-on, its unit is
"kg" and it can be calibrated only under "kg" unit
F1.1=3 "kg" and "lb" unit choosing and converting, whenever turn-on, its unit is
"lb" and it can be calibrated only under "lb" unit
F1.1=4 only for choosing “lb” unit
F1: Function management
F2: General parameters
F3: Calibration
5. Setting and Calibrating
For conveniently setting parameters of function and calibrating, this indicator has
3 main directories on the modification interface.
10