41
To begin:
1. Measure the calibration signal in the power spectrum first.
2. Move the cursor to the calibration frequency. In this example of
acoustic calibration, the calibration frequency is 1 kHz.
3. Open the Calibration table from the [
Tools
] menu. Input the new
calibration values for the calibration signal. In the example of acoustic
calibration, the input value is 94dB, but your value may be different due
to equipment and calibration.
4. Press the [OK] key to leave the calibration table. The new sensitivity
value of each sensor will be assigned to the engineering unit table
automatically.
5. Do the spectrum measurement again to make sure that the calibration
procedure is correct.
Mark cursor
Select Mark Cursor will add a label showing the cursor values. After marking
a cursor, you can move the cursor to other location and mark it again.