BV200 Plasma Viscometer Mk2/3 Customer User
Manual
Issue 9
th
Sep 2020 - Rev 1.8
Page 40 of 53
c. Once complete click the Press here to continue button
After the first stage, the next popup will be displayed, with the following text:-
3.
Empty the decontamination liquids from the waste and condensation bottles then
rinse each with fresh (tap) water very well. Remove the wash tube from the
Decontamination bottle cap tube adaptor and re-connect to the Wash Bottle.
Pour approx. 100ml of tap water into the carriage area near the front of the BV200
unit, so it drips down into the drainage channels below the carriages to flush the
sump valve.
At the last stage, the final popup will be displayed, with the following text:-
4.
Empty the liquids from the Wash, Waste, Decontamination, and Condensation
bottles.
Rinse all four bottles very well with fresh (tap) water.
If required for testing, refill the 2 litre Wash Bottle with the normal amount of wash
fluid
(2 litres of demineralised water and 1 tube of wash additive or bulk additive as
directed)