29
03/ 2020
BENNING MM 10–PV/ MM 10–1
This symbol on the BENNING MM 10-PV/ MM 10-1 indicates that the
BENNING MM 10-PV/ MM 10-1 is equipped with protective insulation
(protection class II).
Please observe the operating manual!
This symbol on the BENNING MM 10-PV/ MM 10-1 means that the
BENNING MM 10-PV/ MM 10-1 complies with the EU directives.
This symbol on the BENNING MM 10-PV/ MM 10-1 indicates the fuses
which it contains.
This symbol appears on the display to indicate a discharged battery.
This symbol indicates the “diode-testing” application.
This symbol designates the „continuity test“ field. The buzzer is intend
-
ed for acoustic result output.
This symbol marks the range “capacity testing”.
(DC) Direct voltage or current.
(AC) Alternating voltage or current
Ground (voltage against ground)
2. Safety instructions
The instrument is built and tested in accordance with
DIN VDE 0411 Part 1/EN 61010-1
DIN VDE 0411 Part 2-033/EN 61010-2-033
DIN VDE 0411 Part 031/EN 61010-031
and has left the factory in perfectly safe technical condition.
To preserve this condition and to ensure safe operation of the device, the user
must observe the notes and warnings given in these instructions at all times.
Improper handling and non-observance of the warnings might involve severe
injuries
or
danger
to life.
WARNING! Be extremely careful when working with bare con-
ductors or main line carrier! Contact with live conductors will
cause an electric shock!
Attention! Magnets might affect the correct functioning of car-
diac pacemakers and implanted defibrillators. As a user of such
medical devices, keep a sufficient distance to the magnet.
The BENNING MM 10-PV/ MM 10-1 must be used in electrical
circuits of overvoltage category III with a conductor for a maxi-
mum of 1000 V to earth or of overvoltage category IV with a
conductor for a maximum of 600 V to earth only.
Only use suitable measuring leads for this. With measurements
within measurement category III or measurement category IV,
the projecting conductive part of a contact tip of the measuring
leads must not be longer than 4 mm.
Prior to carrying out measurements within measurement cat-
egory III and measurement category IV, the push-on caps pro-
vided with the set and marked with CAT III and CAT IV must be
pushed onto the contact tips. The purpose of this measure is
user protection.
Please observe that work on live parts and electrical compo-
nents of all kinds is dangerous! Even low voltages of 30 V AC
and 60 V DC may be dangerous to human life!